Maintenance
Balancing the Sensor
Optics
If the
Service
indicator reappears after a PACER cycle (even after
performing the mirror cleaning procedure above) check the sensor
balance adjustment. Improper adjustment of the optical balance is the
most common cause of instrument malfunction. In addition, new
systems may require an optical balance adjustment after one or two
months of operation.
IMPORTANT:
Always clean the mirror as described above before
performing an optical balance.
Set the sensor mode to HEAT, wait one minute, then observe the
balance indicator. If only the lowest bar is illuminated, or the lower 1/
2 inch of the status bar, the sensor is properly balanced. If not, balance
the sensor according the procedure below. The location of the balance
adjustment screw for various sensors is shown in Figure 7-1 below.
Figure 7-1: Balance Adjustment Screw Locations
Sensor Balance
adjustment screw
Model D2
Model 1111H
Model 1311H
Содержание Optica
Страница 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Страница 11: ...Chapter 1 ...
Страница 20: ...Chapter 2 ...
Страница 43: ...Chapter 3 ...
Страница 58: ...Chapter 4 ...
Страница 80: ...Chapter 5 ...
Страница 81: ...Programming the 4x40 Optica Introduction 5 1 Programming Technique 5 2 Programmable Functions 5 4 ...
Страница 94: ...Chapter 6 ...
Страница 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Страница 99: ...Chapter 7 ...
Страница 109: ...Appendix A ...
Страница 114: ...Appendix B ...
Страница 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 118: ...Humidity Equations and Conversion Chart B 3 January 2006 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 119: ...Appendix C ...
Страница 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing Optica Models D 3 ...
Страница 127: ...Appendix E ...
Страница 128: ...Glossary ...
Страница 130: ...Appendix F ...
Страница 133: ...Appendix G ...
Страница 134: ...Communicating with the OPTICA Using Ethernet Direct Communication G 1 Computer Communication G 1 ...