11-2
CIC Pro™
2026419-033E
Reload software
Purpose
If the CIC Pro center experiences intermittent problems, or the flash drive or hard
drive is replaced, you may need to re-image either the hard drive or flash drive. Refer
to the following table for instructions:
Process overview
This section describes the procedures required to re-image the flash or hard drive:
Gather required equipment and tools on page 11-2
Create a CIC Pro center image restore USB memory stick on page 11-2
Reload CIC Pro center software on page 11-5
Gather required equipment and tools
Service PC (desktop or laptop computer) equipped with a DVD drive, and
running Windows NT, 2000, or XP.
This manual.
CIC v5.1.x Image DVD supplied via the CIC v5.1.x software reload/disaster
recovery kit.
USB memory stick with a minimum of 1 GB memory.
Create a CIC Pro center image restore USB memory
stick
1.
Insert the USB memory stick into a vacant USB port on the service PC.
2.
Insert the DVD containing the
GE Software Distribution Utility
into the service
PC’s DVD drive.
3.
On the
GE Software Distribution Utility
screen, click
List USB Drive(s)
.
If this FRU is replaced
Follow this procedure
Hard drive
Re-image the hard drive. Follow the
Flash drive (solid-state drive)
Re-image the Flash drive. Follow the
.
CPU PCB (mother board)
Do not re-image the Flash drive or hard drive after
replacing the CPU PCB.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......