9-24
CIC Pro™
2026419-033E
Field replaceable units (FRUs)
8.
Re-assemble the cover on page 9-24
9.
Perform the following checkout procedures:
Perform
Ground continuity test on page 10-6
Perform
Ground (earth) wire leakage current test on page 10-7
.
Perform
Enclosure leakage current test on page 10-8
.
Perform general visual inspection.
See Visual inspection on page 10-3.
Inspect fans and power supply.
Inspect connectors.
Inspect cable insulation.
See External surfaces on page 10-4.
Clean internal components.
See Internal components on page 10-5.
Clean fans.
Clean filters.
Check access to all other care units on page 7-7
.
Verify that the power LED is on after the unit is powered up.
Re-assemble the cover
WARNING
SHOCK HAZARD —Whenever the cover is removed, you can
disrupt internal components and potentially disrupt proper power or
ground connections. The potential for electric shock exists. When
the cover is replaced, perform
Electrical safety tests on page 10-6
FRU checkout procedures on page 7-10
Re-assemble the cover on the desktop
1.
Slide the cover over the chassis.
2.
Replace the eight screws that connect the cover to the sides of the chassis.
3.
If you did replace any parts after opening up the cover, you will have to perform
Electrical safety tests on page 10-6
and
FRU checkout procedures on page 7-10
.
Re-assemble the cover on the rack-mount
1.
Slide the cover into the chassis.
2.
Replace the four screws that connect the cover to the back of the chassis.
3.
If you did replace any parts after opening up the cover, you will have to perform
Electrical safety tests on page 10-6
and
FRU checkout procedures on page 7-10
.
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......