Field replaceable units (FRUs)
2026419-033E
CIC Pro™
9-13
WARNING
The CIC Pro center remains energized after AC power is removed.
Wait 30 seconds before proceeding.
4.
Turn off the power to all connected peripheral devices.
5.
Label, then disconnect all peripheral cables and telecommunication lines
connected to the connectors or ports on back of the processor box.
6.
Follow the electrostatic discharge (ESD) precautions on page 9-11
Remove the cover
WARNING
SHOCK HAZARD —Whenever the cover is removed, you can
disrupt internal components and potentially disrupt proper power or
ground connections. The potential for electric shock exists. When
the cover is replaced, perform
Electrical safety tests on page 10-6
FRU checkout procedures on page 7-10
Desktop
1.
Prepare the unit for disassembly on page 9-12
.
2.
Remove the eight screws that connect the cover to the sides of the chassis (retain
the screws).
038
3.
Lift the cover off the chassis.
Rack-mount
1.
Prepare the unit for disassembly on page 9-12
.
2.
Remove the four screws that connect the cover to the back of the chassis (retain
the screws).
Содержание MP100 Series
Страница 12: ...x CIC Pro 2026419 033E...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility...
Страница 313: ......