Measurement & Annotation Setup
1
– Tap the lower-left corner of the display (typically contains a
GE Logo
) or (or press
the
hard key) at any time to open the
Global Menu
, which provides access to
the
Setup Menu
.
2
– Select to alter the
Measurement and Annotation
-specific settings shown here.
3
– View the list of 3DPM Tips already calibrated for use with the attached probe.
4
– Select to view a list of Stereo Tips already-calibrated for use with the attached
probe. The
Add Tip
and
Edit List
options are not intended for general use. Deleting
a tip through the
Edit List
option will permanently remove that tip’s calibration data
requiring the probe and tip to be returned to an authorized GE Inspection Technologies
service center for re-calibration.
5
– Save
Preset
Annotation (notes) to an external storage device or
Import
preset notes
to the instrument from an external device.
6
– Displays or removes the Zoom Window (used for detailed cursor placement during
measurement).
7
– Select
AUTO
to show the 3DPM red/yellow un-measurable pixel mask only when a
cursor is placed on a red/yellow pixel. Select
ON
to always show the pixel mask when
measuring.
8
– Select
Z-Map
for most usage.
P-Map
images contain additional data and may be
requested by GE Technical Support to aid in troubleshooting.
9
– Controls the display of the Stereo Index, which appears during Stereo Measurement.
Click here for more about Stereo Index.
10
– Sets a threshold value for the Stereo Index below which the index flashes during
Stereo Measurement. This may be used to warn the user of low-index measurements
where the accuracy may be reduced.
11
– Specify the unit of measurement as
Inches
or
Millimeters
.