Work with Orders
• Select
View Test
to open the patient test and view it as if the test had been
opened from the
Files
list.
4.
Select
Continue
to detach the order from the existing patient test and attach the
order to the new patient test.
Change the Order Attached to a Patient Test
If an incorrect order is attached to a patient test, use below procedure to detach the
order from the patient test and replace it with a different order.
Before you begin this procedure, make sure that the patient test is not already
transmitted to its default destination. If the test has the
Sent
status of
Yes
(for
example, the test was already sent to its default destination), the order can no longer
be detached from the test. A message displays indicating that the patient test has
been transmitted and the order cannot be detached.
WARNING
:
INACCURATE PATIENT DATA - Incorrect patient information can cause patient
data mismatch. Data assigned to the wrong patient causes erroneous patient
data that can affect diagnosis and treatment.
Be sure to check the patient information screen for each new patient. Make
sure that you attach the correct order for the correct patient.
1.
Detach the incorrect order from the patient test. See
.
2.
Attach the correct order to the patient test. See
Attach an Order when the Patient
Detach an Order from a Patient Test
Before you begin this procedure, make sure that the patient test with the incorrect
order is not already transmitted to its default destination.
NOTE
:
If the patient test to which the order is attached is already transmitted to its
default destination, the order cannot be detached from the test. A message
displays indicating that the patient test has been transmitted and the order
cannot be detached.
1.
From the
Files
list, select the patient report with the incorrect order that you
want to detach.
The patient report displays in the
ECG
tab.
2.
Expand the
Patient Information
screen and scroll down to the
Order Number
field.
3.
Select
Detach
next to the
Order Number
field to clear the field.
82
MAC VU360
™
Resting ECG Analysis System
2088531-370-2