Recording an ECG or Rhythm
If ...
Then ...
Next Steps
Hookup Advisor
status
is yellow or red, there are
issues with the ECG signal
quality during the recording
of this ECG. The
Reject
icon
is highlighted in blue to
reflect the status of
Hookup
Advisor
. The
Accept
icon is
not highlighted.
Select the
Reject
icon:
The ECG patient report
is discarded. The
ECG
Preview
tab closes
and returns to the
Acquisition screen.
Start a new ECG on the same
patient. For more information, see
ECG Acquisition Overview on page
47
.
Review an ECG Patient Report
When you disconnect patient leads and there is no activity with the MAC VU360
device for 2 minutes, all ECG reports will automatically close.
After the 10 seconds ECG is acquired and the ECG preview is accepted, the patient
report is displayed in the
ECG
tab in the configured report format for review.
If the
CRIT
option is enabled on the device, and one or more critical values are
detected during ECG acquisition, a window opens on top of the
ECG
tab displaying
the notifications for the critical values in the order in which they are detected.
Select
Continue
to acknowledge the notification and proceed with other tasks.
54
MAC VU360
™
Resting ECG Analysis System
2088531-370-2