Recording an ECG or Rhythm

These changes apply only to the current patient's test(s). If you start a new

patient, the changes reset to the values configured for the device.

ECG Acquisition Overview

You can record an ECG in pre-acquisition or post-acquisition modes.

Table 15: ECG Acquisition Modes

Acquisition
Mode

Description

Pre-acquisition

When you start recording an ECG:

• If 10 seconds of ECG data is available, the system records the previous 10

seconds of data for analysis.

• If 10 seconds of ECG data is not available, the system continues recording

until 10 seconds of ECG data is obtained.

Post-acquisition When you start recording an ECG, the system records the next 10 seconds of

data for analysis.

If the 

Auto-ECG

 option is enabled on the device, the device automatically records one

ECG for each new patient. For more information on automatic ECG acquisition, see

Automatically Acquire an ECG on page 47

.

To record an ECG patient test manually, see 

Record an ECG Manually on page 48

.

Automatically Acquire an ECG

Start a test for a new patient. For more information, see 

Start a Test for a New Patient

on page 31

.

Make sure that your administrator enables the 

Auto-ECG

 setting.

The system begins to automatically record an ECG:
• When 

Auto-ECG

 setting is enabled.

• If the 

Hookup Advisor

 status is green.

• If you are viewing the scrolling waveforms in the 

Live

 tab.

The automatic acquisition of ECG happens only once per patient connection.
The 

Stop Auto ECG

 icon displays the count of acquisition progress until you record

10 seconds of data. After acquiring 10 seconds of data with good signal quality, the

recording stops and a preview of the ECG patient report displays in the 

ECG

 tab.

If you record the ECG before you enter patient information in the 

Patient Information

screen, you can edit patient information before accepting the preview. For more

information, see 

Accept or Reject an ECG Patient Report on page 50

.

Automatic ECG acquisition is triggered only one time for the current patient test. Start

any new ECG tests for that same patient manually.

2088531-370-2

  MAC VU360

 Resting ECG Analysis System

 47

Содержание MAC VU360

Страница 1: ...Manual Version 1 02 2088531 370 2 Keep this manual with your Regulatory and Safety Manual and equipment at all times and periodically review it MAC VU360 Resting ECG Analysis System Operator s Manual...

Страница 2: ...pment Manufacturer OEM documents go to the device manufacturer s website This document describes the MAC VU360 Resting ECG Analysis System also referred to as the product system or device This documen...

Страница 3: ...ce to obtain the source code of the open source software used in the product if required Support GE Healthcare maintains a trained staff of application and technical experts to answer questions and to...

Страница 4: ...the Device on Standby 27 Perform System Reset 27 3 Patient Information 29 Patient Information Screen Overview 29 Start a Test for a New Patient 31 Enter Patient Information 32 Update Patient Informati...

Страница 5: ...ached to a Different Patient Test 81 Change the Order Attached to a Patient Test 82 Detach an Order from a Patient Test 82 Order Status 83 6 Work with the Files List 85 Review a Stored Patient Report...

Страница 6: ...118 Storage 119 Other Cleaning and Disinfection Agents 119 Additional Information 120 12 Troubleshooting 121 System Errors 121 ECG Acquisition Errors 122 Printing Errors 123 Report Transmission Error...

Страница 7: ...Glossary 167...

Страница 8: ...CG measurement and interpretative analysis prints 12 leads or 15 leads of ECG and transmits ECG data to and from a central ECG cardiovascular information system Front and Rear View Item Part Name Desc...

Страница 9: ...er 6 Front Bin Used for storage 7 Front Wheels and Lock Used to mobilize and lock the trolley in position The front wheels are fixed wheels and non conductive 8 Acquisition Module Cable Connects the A...

Страница 10: ...e focus in a window or to select items from the screen 7 Shift Use to type a capital letter For example press Shift p to type a capital P 8 Space bar Enters a space in the text 9 Alt Switches between...

Страница 11: ...AC Power LED Indicates various AC power states Green light indicates the device is plugged in and receiving power No light indicates the device is not plugged into AC power 15 Power on LED Indicates...

Страница 12: ...d enable the Full Disclosure option 3 Date and Time Current local date and time in the configured date and time format To configure a date and time format refer to the MAC VU360 Resting ECG Analysis S...

Страница 13: ...leges If you disable user authentication or configure with Technician ID access the Default user must to log on as a user with sufficient privileges to access a menu option 9 Start New Patient icon Se...

Страница 14: ...during the recording of a rhythm NOTE A change to the filter speed or gain is applicable to the current patient For a new patient the values are reset to default settings 16 Start Rhythm icon Select...

Страница 15: ...configured with Technician ID access the Default user is prompted to log on as a user with sufficient privileges 3 Service Displays the Service screen used to service the device Your administrator mus...

Страница 16: ...ts The color of the icons change to show the level of battery life You may operate the device connected to the AC Mains power when the batteries are removed The device can also operate with a single b...

Страница 17: ...is critically low and you should connect to AC power immediately No color but crossed out Battery Not Present AC Mains Power Shows that batteryB is not in the device and the AC Mains power is connecte...

Страница 18: ...icon when connected to a LAN or WLAN network Table 4 LAN Icons Network Status Icon Status Description LAN Active The device is connected to a LAN LAN Connected The device is connected to a remote serv...

Страница 19: ...bination works only for English keyboard To increase the brightness of the screen select Alt Arrow up keys on the keyboard To decrease the brightness of the screen select Alt Arrow down keys on the ke...

Страница 20: ...ogged on to the device as the Default User If the user authentication mode is Technician ID enter the Technician ID to log on as a Default User Power Off the ECG Device 1 Before you stop the device co...

Страница 21: ...ser Authentication Mode Description Full authentication with STAT login The Login screen displays with these fields when the device is powered on or unlocked The users below can log on to the device P...

Страница 22: ...default password immediately after the first login Service user The username is Service This username is intended for use by Service personnel A user with the user management privilege can set the pa...

Страница 23: ...e displays on the upper right corner of all of the screens you have access If your login fails see the following table Table 9 Login Errors Symptom Cause Solution You entered your username or password...

Страница 24: ...a valid Technician ID to populate the Technician ID field 2 Select Continue You are logged on as the Default User The Acquisition screen displays You can perform tasks with Default User or Technician...

Страница 25: ...Change Password The Change Password dialog box opens 2 Enter the current password and new password and confirm the new password 3 Select Change Password If the new password meets the password require...

Страница 26: ...t L on the keyboard NOTE This only works when Full Authentication with STAT is enabled The device is locked Your username displays on the lock screen Unlock the Device 1 Perform one of the following s...

Страница 27: ...procedure Unlock the Device on page 26 to unlock and log in to the device If user authentication is not configured you are returned to the last screen displayed when the device was put on standby mod...

Страница 28: ...ges logs and user data will be lost and unrecoverable 2 Enter the serial number of the device in the Enter the system serial number field The Restore button enables when the serial number is entered c...

Страница 29: ...if order management is disabled Read a patient s barcode with a barcode scanner Use a keyboard Do an Admission Discharge Transfer ADT query WARNING INACCURATE PATIENT DATA Patient data can be kept fro...

Страница 30: ...s filled in from the Location ID configured in the System Settings 6 BPM Shows the real time beats per minute BPM The shown heart rate updates whenever the heart rate calculation algorithm reports a c...

Страница 31: ...tion screen 1 Select the Start New Patient icon located in the upper right corner of the Acquisition screen If Then A patient test is open with unsaved data a message tells that unsaved patient data w...

Страница 32: ...nformation You can scan a patient s barcode using an external barcode reader or the internal barcode reader of the Acquisition Module Before you use an external barcode reader make sure that The BRCD...

Страница 33: ...the data scanned from the barcode and existing patient information a message displays Perform one of the steps as follows Select Use Scanned Data to populate the barcode data into the relevant fields...

Страница 34: ...found on the device The orders show on the Orders tab in a filtered list Select the order you will to attach to the patient test One local order is found on the device The order fills in the Patient...

Страница 35: ...mum of five records display on the Acquisition screen Select the ADT record to search the order using that selected record s Patient ID Only one ADT record is found on the MUSE system System searches...

Страница 36: ...the order you will to attach to the patient test One local order is found on the device The order fills in the Patient Information screen if there is no mismatch with the patient data on the screen No...

Страница 37: ...ID order found in that ADT record If multiple orders are found on the system The orders show on the Orders tab in a filtered list Select the order you will to attach to the patient test If only one o...

Страница 38: ...hat no matching remote orders are found If remote order query is failed on the system A message displays in the notification area that remote order query has failed No ADT records are found on the MUS...

Страница 39: ...uery is done only on Site 1 if the MUSE system version is MUSE v9 SP5 or earlier ADT query to the MUSE system for sites other than Site 1 requires MUSE v9 SP6 or later Enter or Edit Patient Informatio...

Страница 40: ...your entries select Save The information is saved and the Patient Information screen closes Based on the Mandatory fields apply for Transmission or Acquisition settings the ECG report will not be acce...

Страница 41: ...save time and prevent the necessary to restart the ECG The Hookup Advisor reports the status based on signals from each leadwire When you connect the RA R leadwire and another leadwire to the patient...

Страница 42: ...e does not automatically expand when the overall Hookup Advisor status is yellow or red Only the Hookup Advisor lead quality status indicator and status messages display in the Notification Area 3 Lea...

Страница 43: ...dicator and contains a checkmark The two circles to the left are black A message displays showing that the signal lead quality is good to acquire an ECG NOTE The background color of the Start ECG icon...

Страница 44: ...isition you can accept the ECG If the quality of the ECG signal has problems during the 10 seconds ECG acquisition the status changes from green to yellow or red It is recommended to reject an ECG tha...

Страница 45: ...to collapse it These changes apply only to the current patient test s If you start a new patient the changes reset to the values configured for the device Enable ACS Interpretation You can enable the...

Страница 46: ...ACS algorithm is not executed Change Speed Gain and Filter The Acquisition screen displays the waveform based on the configured speed gain and filter After starting a new patient you can change the sp...

Страница 47: ...ecord an ECG Manually on page 48 Automatically Acquire an ECG Start a test for a new patient For more information see Start a Test for a New Patient on page 31 Make sure that your administrator enable...

Страница 48: ...stops The automatic ECG function is aborted and the device will function in manual ECG mode if An automatic ECG acquisition is stopped before 10 seconds of data is acquired The preview of an ECG that...

Страница 49: ...of ECG data for analysis The Start ECG icon changes to Stop ECG and the 10 seconds count of acquisition progress is displayed on the icon You can cancel the acquisition before 10 seconds of data is re...

Страница 50: ...smission queue until you enter the patient demographic data for the mandatory fields An error message displays in the notification area You need to complete the data for the mandatory fields and manua...

Страница 51: ...enable the Full Disclosure option You can accept this preview to save the ECG patient report in the Files tab or reject it and start another ECG Table 16 ECG Tab Displaying a Preview of the ECG Patien...

Страница 52: ...go back to the preview and accept or reject it 6 Full Disclosure tab Click the Full Disclosure tab from the patient record menu bar to navigate to the Full Disclosure waveform and application The ECG...

Страница 53: ...ueue of pending reports to be sent to the configured automatic destination For more information see Display the Report Queue on page 99 NOTE Based on the Mandatory fields apply for Transmission settin...

Страница 54: ...view an ECG Patient Report When you disconnect patient leads and there is no activity with the MAC VU360 device for 2 minutes all ECG reports will automatically close After the 10 seconds ECG is acqui...

Страница 55: ...more information on the report formats and the standard layout of the ECG patient report see ECG Report Formats on page 130 3 Gain and Filter To change the waveform gain or filter select anywhere arou...

Страница 56: ...nt on the MAC VU360 device To start an ECG for a new patient select Start New Patient For more information see Start a Test for a New Patient on page 31 To start a new 10 seconds ECG on the same patie...

Страница 57: ...ing print jobs are cancelled Record a Rhythm Ensure that sufficient paper is available in the paper tray to print a rhythm report A rhythm report can be stored in digital form in the Files list or pri...

Страница 58: ...ist The rhythm is not printed Go to step 4 if you want to stop recording the rhythm before its configured duration is reached otherwise go to step 5 Both The rhythm for the patient is digitally record...

Страница 59: ...cription 1 Patient Information Banner Displays patient information such as patient first name last name gender age and so on Select anywhere in the Patient Information banner to edit patient informati...

Страница 60: ...b Select the Live tab to return to the live waveform display in the Acquisition screen To start a new rhythm for the same patient select the Live tab to return to the live waveform display in the Acqu...

Страница 61: ...ation with folder path PDF format Full Disclosure USB R W flash drive PDF format Full Disclosure DCP server destination MUSE v9 SP6 or higher or MUSE NX and MUSE DICOM Gateway Pro SP6 or higher PDF fo...

Страница 62: ...ys on the notification bar in the lower right side of the screen in the format Destination_Name Job_Status For example if the destination name is USB and the job status is Failed the status displays a...

Страница 63: ...u enter the patient demographic data for the mandatory fields If you manually transmit ECG report an error message Unable to transmit Incomplete patient data displays on the Acquisition screen You nee...

Страница 64: ...ythm in the selected report format is sent to the printer The patient report is printed in the order in which it was received If no other patient reports are printing the report is printed immediately...

Страница 65: ...DATA Incorrect patient information can cause patient data mismatch Data assigned to the wrong patient causes erroneous patient data that can affect diagnosis and treatment Be sure to check the patient...

Страница 66: ...on message may display if you are trying to delete a patient report that has not yet been transmitted to the default destination if a message has been configured by your administrator Perform one of t...

Страница 67: ...e Overview The Full Disclosure ECG option shows one lead of the patient waveform for a maximum of 5 minutes From this waveform you can create a Full Disclosure report FD Report or create a 12 lead or...

Страница 68: ...creen displays the previous 5 minutes of Full Disclosure ECG data If you disconnect the acquisition module the ECG acquisition stops in 10 seconds 3 To change the lead click Select Lead All the config...

Страница 69: ...atient Report see Accept or Reject an ECG Patient Report on page 50 d To review an ECG Patient Report see Review an ECG Patient Report on page 54 6 To generate a Full Disclosure report click FD Report...

Страница 70: ...left and right arrows on the screen to see the next page When you record Full Disclosure ECG If the acquisition module is disconnected the Full Disclosure report displays empty space on the screen If...

Страница 71: ...formation for the Full Disclosure report For more information see Edit Patient Information in a Patient Report on page 65 7 Copy icon Prints a copy of the Full Disclosure report For more information s...

Страница 72: ...You can retrieve orders from an order management server such as a MUSE system that is connected to your network When the orders list is updated either automatically or manually new orders are populate...

Страница 73: ...or Visit Number 4 Navigation arrows Navigates to the previous and next pages in the Orders list 5 Last Updated date and time Displays the date and time the order list was last updated 6 Refresh icon...

Страница 74: ...Only one order can be associated with a patient test at any given time Orders cannot be attached to Transmitted patient reports or Digital Rhythm patient reports If you do not have permissions to edi...

Страница 75: ...ually update the Orders list Manually Update the Orders List You can manually update the orders list at any time even if the auto update option is enabled 1 From the Acquisition screen select the Orde...

Страница 76: ...SAP Pre Op Call back Routine If you select the Priority column again the sorting order is reversed You can change the sort order by selecting any one of the column headers The orders list is sorted in...

Страница 77: ...ocation filter in the Orders list the filter icon changes to to indicate that the order list is filtered and does not display all orders If you select Then A pre configured filter group The orders lis...

Страница 78: ...the device configured in the device settings 3 To display an expanded list of Orders select the Expand icon The Orders expanded list opens By default the applied filter is Current Patient Location and...

Страница 79: ...ation and select Save to save the patient data 4 Record the ECG For more information see Record an ECG Manually on page 48 5 Verify that the status of the order in the Orders list is Attached Attach a...

Страница 80: ...patient report All patient demographic fields included in the order are populated in the test overwriting the existing patient data The status of the order is changed to Attached 4 Update test demogr...

Страница 81: ...nformation Using the Keyboard on page 39 Attach an Order that is Attached to a Different Patient Test NOTE Only one patient test can be associated with an order at any given time regardless of the sta...

Страница 82: ...that can affect diagnosis and treatment Be sure to check the patient information screen for each new patient Make sure that you attach the correct order for the correct patient 1 Detach the incorrect...

Страница 83: ...der to a patient test from the Orders list The order status changes from Open to Attached The MUSE server is notified to change the status of the corresponding order on the MUSE system from Open to Pe...

Страница 84: ...Work with Orders NOTE The Attached order status is not included in the transmitted patient reports 84 MAC VU360 Resting ECG Analysis System 2088531 370 2...

Страница 85: ...collapsed list Table 22 Files Collapsed List Item Name Description 1 Files tab Opens the Files collapsed list that stores the saved patient reports 2 Expand icon Opens the Files expanded list 3 Files...

Страница 86: ...ient reports 2 Print List Prints the stored records from the Files list This button is enabled only if the stored records are available 3 Collapse icon Collapses the Files list 4 Files expanded list c...

Страница 87: ...you can only view the patient reports you created in the current session 1 From the Acquisition screen select the Files tab The Files collapsed list opens 2 Select the Rhythm or Full Disclosure or EC...

Страница 88: ...more information see Delete Stored Patient Reports from the Files List on page 95 5 Edit icon Select the Edit icon to expand the Patient Information screen and edit patient information for the patien...

Страница 89: ...ing ECG Shared directory destination with folder path PDF Hilltop and Sapphire XML Sapphire XML format is available if you enable the XML format output in the Option Manager formats Digital Rhythm USB...

Страница 90: ...To another configured destination 1 Select anywhere around the ellipsis icon on the lower left side of the screen to expand the Transmit menu 2 From the expanded Transmit menu select any configured d...

Страница 91: ...on a message displays in the notification asking you to confirm if you want to re transmit the already transmitted reports Perform one of the actions below Select OK to re transmit the patient report...

Страница 92: ...h you want to print a copy The selected patient report opens in a new tab An ECG patient report displays in the ECG tab A Rhythm report displays in the Rhythm tab A Full Disclosure report displays in...

Страница 93: ...quisition settings the ECG report is not accepted transmitted or printed until you enter the patient demographic data for the mandatory fields If you manually print an error message Unable to print In...

Страница 94: ...on the Acquisition Module Edit Patient Information in a Stored Patient Report Make sure that you have the privilege to open stored rhythm Full Disclosure or ECG patient reports from the Files list and...

Страница 95: ...ns with a grey background indicating that this is a stored patient report 3 Edit the patient information using a keyboard See Enter or Edit Patient Information Using the Keyboard on page 39 4 Select S...

Страница 96: ...e the patient reports The selected patient reports are deleted from the Files list An alert may be configured by your administrator to warn you before deletion of untransmitted reports If this alert i...

Страница 97: ...transmitting to a designated location are temporarily stored in the Queue list The successfully transmitted digital rhythm FD report and ECG patient reports are immediately deleted from the Queue lis...

Страница 98: ...e figure illustrates the Queue expanded list Table 26 Queue Expanded List Item Name Description 1 Queue tab Displays the list of patient reports in the transmission queue 2 Collapse icon Collapses the...

Страница 99: ...lays the re tried transmission count in numbers If the device sends the report on the first attempt the Retry Count is 0 Job Failure Reason Displays the reason for the failed transmission If the devic...

Страница 100: ...and then try again if necessary If the job is in Not Sent or Failed status a message displays asking you to confirm the deletion of the selected job 5 Select Delete to confirm the deletion The select...

Страница 101: ...gins transmission as soon as the current patient report is transmitted If a patient report is successfully transmitted it is immediately deleted from the Queue Review the queue to confirm that the pat...

Страница 102: ...ys in the Acquisition screen The list of recent patients is stored in the Patients list and you can view up to 500 recent patient records The following figure illustrates the Patients collapsed list T...

Страница 103: ...psed list Patient ID Last Name First Name Gender Displays the following six columns in the expanded list Patient ID Last Name First Name DOB Gender Race By default the list is sorted by Last Name in a...

Страница 104: ...tion see Start a Test for a New Patient on page 31 2 Select the Patients tab on the right side of the Acquisition screen The Patients collapsed list opens and displays a list of patients 3 Double tap...

Страница 105: ...remote applications and you can open a maximum of three tabs at a time If the remote application is slow to load the following message displays Loading URL Any errors generated because of the link dis...

Страница 106: ...105 2 Open the PDF To navigate the PDF do any of the following Move through the PDF Select to move to the next page in the PDF Select to move to the previous page in the PDF Enter a specific page numb...

Страница 107: ...different paper size The paper tray is configured to use the appropriate paper size for the destination location Use the instructions to change the paper size by moving the spacers in the paper tray 1...

Страница 108: ...ottom of the tray and the tabs on both sides of the spacer are affixed to the designated tab spaces on the tray 5 Push the paper tray back to the printer assembly Store Thermal Paper When imaged and s...

Страница 109: ...opes or sheet separators made of polystyrene polypropylene or polyethylene do not degrade thermal traces However these materials afford no protection against fading from external causes Do NOT store t...

Страница 110: ...me To make sure a fully charged battery in between acquisitions power off the system and connect it to an AC wall outlet until you use the system again This prolongs battery runtime The battery status...

Страница 111: ...ect the system to an AC wall outlet 3 Charge the battery for 3 hours or until the battery status indicator displays a full charge Replace the Battery NOTICE BATTERY PACK DISPOSAL Do not dispose of the...

Страница 112: ...ries into the battery compartment slots in the correct orientation and direction indicated by the arrow 4 Lift the battery compartment door in the upward direction and close it 112 MAC VU360 Resting E...

Страница 113: ...tions that contain wax or wax compounds to clean the equipment Do not immerse the device in any liquid as this may corrode metal contacts and affect signal quality Do not drip or expose the writer ass...

Страница 114: ...ce Manual Contact your local GE Healthcare service representative for further information For instruments produced by another manufacturer reference the manufacturer s instructions for use Visual Insp...

Страница 115: ...or if you suspect a problem Check the pins that the leadwires plug into They should not be bent or loose Contact GE Healthcare Customer Support for any repairs needed Check the Acquisition Module plas...

Страница 116: ...ion Module Ensure that thorough cleaning is performed to remove all visible soil and organic matter prior to disinfection See Cleaning the Acquisition Module on page 115 1 Dispense fresh Super Sani Cl...

Страница 117: ...astor wheels are functioning properly If you notice any items that need repair contact an authorized service representative to make the repairs Discontinue using the device until the appropriate repai...

Страница 118: ...ley on page 117 1 Dispense fresh Super Sani Cloth Wipe s from the canister 2 Apply disinfectant to the entire surface using fresh wipes Treated surfaces must remain visibly wet for a minimum of three...

Страница 119: ...es is the recommended cleaning and disinfection solution which has been validated on the device However below products are compatible with the device and may be used for cleaning and disinfection It s...

Страница 120: ...using appropriate equipment materials and personnel and achieves the desired result This requires validation and routine monitoring of the process Any deviation from the provided recommendations shoul...

Страница 121: ...ist to create additional storage space Memory is full This ECG cannot be saved The device storage capacity is full Transmit patient reports to configured destinations and delete the transmitted report...

Страница 122: ...lure The keyboard is not working Contact your GE Healthcare Service Support representative if the error persists Touchscreen failure The touchscreen is not working Contact your GE Healthcare Service S...

Страница 123: ...If the error persists contact your GE Healthcare Service Support representative Printing Errors Table 29 Printer Errors Encountered During ECG Patient Report Printing Error Message Error Condition Ho...

Страница 124: ...hm printing Table 31 Printing Errors Encountered During Stored Records List Printing Error Message Error Condition How to troubleshoot Printer error Paper jam detected Paper is jammed 1 Carefully remo...

Страница 125: ...ror persists contact your GE Healthcare Service Support representative Printer error Restart system Contact Service Unknown error or hardware failure in printer Restart the system to check if the erro...

Страница 126: ...um limit of 1 000 reports Wait for the reports in the queue to transmit and try again Unable to transmit Incomplete patient data Unable to transmit one or more reports Incomplete patient data One or m...

Страница 127: ...ive Make sure that the USB flash drive is firmly inserted into the USB port Hilltop generation unsuccessful PDF generation unsuccessful Sapphire generation unsuccessful The file generation is unsucces...

Страница 128: ...client certificate Invalid Client Private Key Password The client private key password is invalid Enter a valid client private key password Client Certificate is invalid The client certificate is inv...

Страница 129: ...Incompatible firmware version of FIPS module Contact your GE Healthcare Service representative to replace the FIPS module with a compatible version FIPS initialization failed Reconfigure wireless If p...

Страница 130: ...wo columns of six rows of waveforms Each column is 5 seconds wide and printed at 50 mm s One column is printed on each page This results in a two page report Multiple page report 2 by 5s 1 Rhythm Ld S...

Страница 131: ...rt Formats Report Format Name Description One Page or Multiple Page Report Computer Graphic Record CGR Shows three columns of four rows of medians at 25 mm s on the left side and three rows of wavefor...

Страница 132: ...econds wide and printed at 25 mm s The lower part isone row of 10 seconds of one lead The rhythm lead shown in the report is configurable One page report 5 by 2s 3 Rhythm Lds Shows two parts The upper...

Страница 133: ...rt One page report Expanded Median Shows the following parts The upper part of the report is text including the measurements but not the interpretation and the measurement matrix which is the measurem...

Страница 134: ...ntal frontal and sagittal planes There are four charts for each plane showing the entire cardiac cycle at the selected gain The gain can be selected from 20 mm mv 40 mm mV 80 mm V or 160 mm mV Each ch...

Страница 135: ...ut the patient such as First Name and Last Name Age and or Date of Birth Gender Race Height and Weight Room Location 2 Vital Signs Displays information about the patient s vital signs such as Heart ra...

Страница 136: ...CG patient report in the configured report format for the selected lead set 8 Pace Spikes Displays pace spikes for patients with a pacemaker The pace spikes represent pacemaker pulses NOTE When two pa...

Страница 137: ...y Visit Number Medical History List of Medications Rhythm Report Format Rhythm reports contain patient information waveform data and ECG acquisition data A rhythm report is a continuous recording of a...

Страница 138: ...tient If the report contains more than one page the BPM changes on each page of the waveform data 5 The waveform data The rhythm report contains waveform data for 12 leads configured for the rhythm re...

Страница 139: ...11 The Gain of the ECG waveform measured in mm mV NOTE You can change the gain before or during the recording and or printing of a rhythm For the printed rhythm report the printing stops and restarts...

Страница 140: ...25 mm s 1 second of data is shown in 25 mm 5 large squares on the rhythm report When recording 10 mm mV 1 mV of data is shown in 10 mm mV 2 large squares on the printout Full Disclosure FD Report Form...

Страница 141: ...shown on the printed or transmitted FD report when the acquisition module is disconnected NOTE A square waves is shown on the printed or transmitted FD report when the lead is disconnected 5 Prints th...

Страница 142: ...e lead that tells the gain on the printed FD report For the digital FD report the calibration pulse of the last gain selected during recording is shown on the full report 11 Prints the Speed of the EC...

Страница 143: ...ree of excessive hair NOTE Do not use solvents to clean the skin solvents trapped under electrodes may lead to abnormal skin reactions 3 Apply the electrodes to the prepared sites WARNING ELECTRIC SHO...

Страница 144: ...ng print jobs in the queue the Patient Information screen is expanded You can enter patient data for the new patient If there are pending print jobs from the previous patient test in the queue a messa...

Страница 145: ...erior axillary line on the same horizontal level as 4 6 V6 violet C6 violet Mid axillary line on the same horizontal level as 4 and 5 7 LA black L yellow Left deltoid 8 LL red F green Above the left a...

Страница 146: ...C6 violet Mid axillary line on the same horizontal level as 4 and 5 7 LA black L yellow Left deltoid 8 LL red F green Above the left ankle alternate placement upper leg as close to the torso as possib...

Страница 147: ...al level as 4 and 5 7 LA black L yellow Left deltoid 8 LL red F green Above the left ankle alternate placement upper leg as close to the torso as possible 9 RL green N black Above the right ankle alte...

Страница 148: ...ectrode Placement To acquire a Vector Loops ECG use the electrode placement shown in the following diagram Table 50 Vector Loops Electrode Placement Item AHA Label IEC Label Description 1 LA black L y...

Страница 149: ...ine on the same horizontal level as 3 and 4 8 RA white R red Just below the clavicle of the right arm 9 H orange H light blue Back of the neck avoid the carotid artery and jugular vein 10 M orange M l...

Страница 150: ...Fields Patient Information Text Box Names When an order is attached to a patient test all Patient Information text boxes are read only The text box names in the table with an asterisk are not 150 MAC...

Страница 151: ...or 9 if the patient year of birth is between 2000 to existent year xx patient place of birth g patient gender male odd number female even number Patient ID Identification number given to the patient...

Страница 152: ...or in the long formats as follows yyyymmdd xxgc yyyymmdd xxgc yyyymmddxxgc yy and yyyy patient year of birth dd patient day of birth mm patient month of birth or patient age NOTE If or is not before t...

Страница 153: ...of birth is between 2000 to current year g patient gender male odd number female even number xx patient place of birth Mandatory fields apply for The mandatory fields that can be configured for Transm...

Страница 154: ...o 232 cm Weight Patient weight in pounds lb or kilograms kg refer to the configured unit of measurement Maximum 3 characters 0 to 999 lb 0 to 454 kg Gender Patient gender Not Applicable Male Female Da...

Страница 155: ...c American Indian Eskimo Hawaiian Pacific Islander Asian Unknown Other Order Number Unique order number given to a patient test If the order number is given by the computer when an order is attached t...

Страница 156: ...acters are supported Bed Number The bed number where the ECG is to be done Maximum 32 characters A to Z a to z 0 to 9 All characters are supported Priority Priority of ECG patient test Not Applicable...

Страница 157: ...All characters are supported Referring MD ID Referring physician ID Maximum 5 characters 0 to 65534 Technician Technician doing the ECG If you are a local user and your user account is configured wit...

Страница 158: ...o z 0 to 9 All characters are supported Number 10 characters Extra optional one special character or in the beginning 0 to 9 Yes or No or Unknown Question 1 Question 2 Question 3 Question 4 The name o...

Страница 159: ...Values Attending MD First Name Attending physician first name Maximum 20 characters A to Z a to z 0 to 9 All characters are supported Attending MD ID Attending physician ID Maximum 5 characters 0 to 6...

Страница 160: ...m to 62 in 158 cm Width 19 in 48 cm Depth 28 in 71 cm System with fixed height premium trolley Handle height 35 in 89 cm or 37 in 94 cm Maximum height 56 in 142 cm to 58 in 147 cm Width 19 in 48 cm De...

Страница 161: ...le 54 Keypad Specifications Item Description Type Sealed elastomer membrane keyboard with tactile feedback Keys Function keys Alphanumeric keys Writer controls Trimpad cursor controls Keypad Keypad an...

Страница 162: ...rge when device is off standby Battery capacity The system shall be able to continuously display scrolling waveforms without printing for a minimum of six hours when the maximum total battery capacity...

Страница 163: ...g Specifications Item Description ECG Interpretation Marquette 12SL ECG Analysis Program for Adults and Pediatrics Computerized measurements 15 lead analysis includes measurements of user selectable a...

Страница 164: ...ications Item Description Operating Conditions Temperature 10 C to 40 C 50 to 104 F Relative Humidity RH 20 to 95 non condensing Atmospheric Pressure 70 to 106kPa Transport Storage Conditions Temperat...

Страница 165: ...ype CF defibrillation proof applied part Network Specifications Table 63 Network Specifications Item Description Frequency bands of transmission 2 401 2 461 GHz 5 180 5 825 GHz Maximum radiated power...

Страница 166: ...non FIPS PSK PEAP MSCHAPV2 PEAP GTC EAP TLS TTLS MSCHAPV2 TTLS GTC Authentication Methods FIPS PSK PEAP GTC EAP TLS Wired Specifications Ethernet interface 802 3 Ethernet interface via RJ45 connector...

Страница 167: ...e amplitude of the waveforms A higher gain makes the amplitude of the waveform appear higher a lower gain makes the amplitude of the waveform appear lower The 10 5 mm mV setting is used to display the...

Страница 168: ...53223 USA 1 414 355 5000 1 800 558 7044 US Only GE Medical Systems Information Technologies GmbH Munzinger Stra e 5 79111 Freiburg Germany 49 761 45 43 0 www gehealthcare com GE Medical Systems Infor...

Отзывы: