Patient Preparation
Table 49: NEHB Electrode Placement
Item AHA Label
IEC Label
Description
1
A1 orange
Nst white
Attachment point is the second rib of the right sternal edge.
2
A2 orange
Nax white
Fifth intercostal space on the left posterior axillary line (same
position as V7 or C7).
3
V4 blue
Nap white
Mid-clavicular line in the fifth intercostal space (same as C4).
Vector Loops Electrode Placement
To acquire a Vector Loops ECG, use the electrode placement shown in the following
diagram.
Table 50: Vector Loops Electrode Placement
Item AHA Label
IEC Label
Description
1
LA black
L yellow
Just below the clavicle of the left arm.
2
E orange
E light blue
Mid-sternum on the same horizontal level as 3 and 4.
3
V4 blue
C4 brown
Mid-clavicular line in the fifth intercostal space.
4
V6 violet
C6 violet
Mid-axillary line on the same horizontal level as 3.
5
LL red
F green
Left leg, in the lower abdominal quadrant.