4-16
MAC™ 800
2031504-159D
Maintenance
Replacing the Mainboard Assembly
Processing ECGs in Internal Storage
If the system has the internal storage option, process any ECGs
remaining in storage by transmitting to your archival system and/or
print them to ensure you have a printed record before proceeding with
the mainboard replacement.
Saving System Configuration Settings
1. Store the
System Configuration
settings to an SD card.
a. Insert SD card in the SD card slot.
b. From the
Main Menu
, press
F4
to select
System Configuration
.
c. Press
F6
(
More
) >
F6
(
More
) >
F3
to select
Export Setup
.
d. Highlight the system setup file you want to export to the SD
card.
e. Press
F1
to select
Export
.
f. When the
Configuration was successfully exported
message is
displayed, press
F6
to select
OK
.
g. Remove the SD card and store in a secure location.
2. Print the
System Setup Report
if you feel you may need it for
additional reference after the FRU replacement procedure.
a. From the
Main Menu
, press
F4
to select
System Configuration
.
b. Press
F6
(
More
) >
F3
to select
Print Setup Report
.
c. Move the focus to the
Complete Setup
button and press
Enter
.
d. Save the printed setup report in a secure location. It can be used
as a reference if
System Setup
needs to be restored manually.
Removing the Mainboard Assembly
1. Disconnect the system from AC power.
2. Remove the battery assembly as described in
3. Remove the top cover assembly as described in
4. Remove the LCD assembly as described in
5. Disconnect the printer cable from the mainboard.
Содержание MAC 800
Страница 7: ...2031504 159D MAC 800 1 1 1 Introduction ...
Страница 13: ...2031504 159D MAC 800 2 1 2 Equipment Overview ...
Страница 18: ...2 6 MAC 800 2031504 159D Equipment Overview Detailed Description Block Diagram ...
Страница 19: ...Equipment Overview 2031504 159D MAC 800 2 7 System Architecture Overview ...
Страница 21: ...Equipment Overview 2031504 159D MAC 800 2 9 Software Architecture Layered Structure of application software ...
Страница 22: ...2 10 MAC 800 2031504 159D Equipment Overview ECG Data Flow With Sampling Rates ...
Страница 23: ...2031504 159D MAC 800 3 1 3 Troubleshooting ...
Страница 50: ...3 28 MAC 800 2031504 159D Troubleshooting ...
Страница 51: ...2031504 159D MAC 800 4 1 4 Maintenance ...
Страница 82: ...4 32 MAC 800 2031504 159D Maintenance ...
Страница 83: ...2031504 159D MAC 800 5 1 5 Parts Lists ...
Страница 86: ...5 4 MAC 800 2031504 159D Parts Lists ...
Страница 87: ...Parts Lists 2031504 159D MAC 800 5 5 ...
Страница 88: ...5 6 MAC 800 2031504 159D Parts Lists ...
Страница 89: ...Parts Lists 2031504 159D MAC 800 5 7 ...
Страница 90: ...5 8 MAC 800 2031504 159D Parts Lists ...
Страница 103: ...2031504 159D MAC 800 A 1 A Technical Specifications ...
Страница 108: ...A 6 MAC 800 2031504 159D Technical Specifications ...
Страница 109: ...2031504 159D MAC 800 B 1 B Electromagnetic Compatibility ...
Страница 115: ...2031504 159D MAC 800 C 1 C Supplies Accessories ...
Страница 121: ......