Revision G
MAC
TM
600
4-11
2047426-002
Maintenance:
Preparing System for FRU Replacement
Reworking the Power Supply (Class II to Class I)
The MAC600 Class II Power Supply (PS) FRU P/N 2047328-001 is
obsolete. The new power supply is qualified as Class I and the P/N is
2047328-002. In order to use the Class I power supply in the MAC600,
the safety classification of MAC600 device needs to change from Class II
to Class I .
Depending on the Serial number of the device different levels of rework
are required. See the following table for further details.
NOTE
All MAC 600 shipments to China are Class I. Therefore, further
rework is not required. For China, see
Supply Medical Class I (For China Only)”
Table 4-6. Power Supply Labels
Legend
Description
1
Class II Power Supply label
2
Class I Power Supply label
If Serial Number is:
Then Perform:
less than or equal to
SF713346059PA
1.
3. Change the Power Supply FRU (P/N
2047328-002).
between SF713346060PA
and SF713496708PA
1.
2. Change the Power Supply FRU (P/N
2047328-002).
equal to or greater than
SF713500001PA
1. Change the Power Supply FRU (P/N
2047328-002).
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...