5-38
MAC™ 3500 Resting ECG Analysis System
Revision L
2021337-036
Maintenance:
Functional Checkout Procedures
Electrical Safety Checks
20. Current leakage and ground continuity test results meet requirements?
Perform electrical safety checks when indicated. All indicated electrical safety
checks require a pass/fail indication for steps performed. Record the
measurement values in your debrief.
Table 6. Electrical Safety Checks
Step
1
Condition
2
UUT - ON
Result
Leakage
Current Limits
Earth Leakage Current
1.
Forward Polarity
NC
______
Pass/Fail
300
2.
Neutral open, Forward Polarity
SFC
______
Pass/Fail
1,000
3.
Neutral open, Reverse Polarity
SFC
______
Pass/Fail
1,000
4.
Reverse Polarity
NC
______
Pass/Fail
300
Enclosure Leakage Current
1.
Forward Polarity
NC
______
Pass/Fail
100
2.
Neutral open, Forward Polarity
SFC
______
Pass/Fail
300
3.
Ground open, Forward Polarity
SFC
______
Pass/Fail
300
4.
Ground open, Reverse Polarity
SFC
______
Pass/Fail
300
5.
Neutral open, Reverse Polarity
SFC
______
Pass/Fail
300
6.
Reverse Polarity
NC
______
Pass/Fail
100
Patient Leakage Current to Ground
1.
Forward Polarity
NC
______
Pass/Fail
10
2.
Neutral open, Forward Polarity
SFC
______
Pass/Fail
50
3.
Ground open, Forward Polarity
SFC
______
Pass/Fail
50
4.
Ground open, Reverse Polarity
SFC
______
Pass/Fail
50
5.
Neutral open, Reverse Polarity
SFC
______
Pass/Fail
50
6.
Reverse Polarity
NC
______
Pass/Fail
10
Patient Leakage Current Mains on Applied Part
3
1.
Forward Polarity
Neutral / Ground Closed
SFC
______
Pass/Fail
5000
2.
Reverse Polarity
Neutral / Ground Closed
SFC
______
Pass/Fail
5000
Содержание MAC 3500
Страница 1: ...GE Healthcare MACTM 3500 Resting ECG Analysis System Version 9D Service Manual 2021337 036 Revision L ...
Страница 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Страница 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Страница 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Страница 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 40: ...3 16 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Installation General Assembly For your notes ...
Страница 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Страница 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 44: ...4 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Troubleshooting Assembly Descriptions ...
Страница 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Страница 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 122: ...5 40 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Maintenance Functional Checkout Procedures ...
Страница 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Страница 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 152: ...6 30 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Parts Lists Field Replaceable Units FRUs ...
Страница 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Страница 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 160: ...A 8 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Technical Specifications Barcode Scanner For your notes ...
Страница 161: ...Revision L MAC 3500 Resting ECG Analysis System B 1 2021337 036 B Software Hardware Compatibility ...
Страница 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 167: ...Revision L MAC 3500 Resting ECG Analysis System C 1 2021337 036 C Electromagnetic Compatibility ...
Страница 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Страница 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Страница 179: ......