Facility needs
LOGIQ V3/V5/V5 Expert – Basic Service Manual
2-15
5726264-100 English Rev.8
2-3-3 Required facility needs(continued)
NOTE:
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound unit is only a conduit.
•
Dedicated single branch power outlet of adequate
amperage meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the unit’s proposed
location
•
Door opening is at least 76 cm (30 in) wide
•
Proposed location for unit is at least 0.5m (1.5 ft.) from the
wall for cooling
•
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
unit to connect cables.
•
Power outlets for other medical equipment.
•
Power outlets for test equipment within 1 m (3.2 ft.) of
Ultrasound system.
•
Clean and protected space to store probes (in their cases or
on a rack)
•
Material to safely clean probes (done with a plastic
container, never metal)
For the amperage requirements, see:
2-3-4 Desirable features
•
Door is at least 92 cm (3 ft.) wide
•
Circuit breaker for dedicated power outlet is easily
accessible
•
Sink with hot and cold water
•
Receptacle for bio–hazardous waste, like used probe
sheaths
• Emergency
oxygen
supply
•
Storage for linens and equipment
•
Nearby waiting room, lavatory, and dressing room
•
Dual level lighting (bright and dim)
•
Lockable cabinet ordered by GE for its software and
proprietary manuals
Содержание LOGIQ V3
Страница 5: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 3 5726264 100 English Rev 8...
Страница 6: ...i 4 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 7: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 5 5726264 100 English Rev 8...
Страница 8: ...i 6 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 9: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 7 5726264 100 English Rev 8...
Страница 10: ...i 8 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 11: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 9 5726264 100 English Rev 8...
Страница 12: ...i 10 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 13: ...LOGIQ V3 V5 V5 Expert Basic Service Manual i 11 5726264 100 English Rev 8...
Страница 18: ...i 16 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 56: ...Introduction 1 32 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 124: ...System Setup 3 48 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 180: ...Components and Functions Theory 5 6 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 222: ...Diagnostics Troubleshooting 7 34 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 252: ...Replacement Procedures 8 30 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 264: ...Renewal Parts 9 12 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 300: ...Care and Maintenance 10 36 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 304: ...Index 4 LOGIQ V3 V5 V5 Expert Basic Service Manual 5726264 100 English Rev 8...
Страница 305: ...GE...