Basic Service Manual
Section 8-6 - Replacing Monitor and Monitor Arm
8 - 77
8-6-3
Replacement of the LCD ARM-T1
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-6-3-1
Preparations
1.) Remove the Ultrasound Bezelless 23.8inch LCD T2 as described in
8-6-6 "Replacement of the Main
ULTRASOUND BEZELLESS 23.8INCH LCD T2" on page 8-84
2.) Remove the T1 Monitor Arm Cover Set as described in
8-6-2 "Replacement of the T1 Monitor Arm
8-6-3-2
Removal Procedure
Table 8-33 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 10 minutes
Phillips Screwdriver
Nipper
Hexagonal (Allen) wrench
Cut resistant gloves
Step
Corresponding graphic
1. Cut 6 cable ties on the left, right, and rear side of the
LCD ARM-T1.
Содержание LOGIQ Totus
Страница 2: ......
Страница 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Страница 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Страница 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Страница 198: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 4 32 Section 4 4 Site Log...
Страница 236: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual 5 38 Section 5 14 Product Manuals...
Страница 627: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual Chapter 10 Care Maintenance 10 25...
Страница 642: ...GE HealthCare CONFIDENTIAL Direction 5936427 Revision Rev 1 LOGIQ Totus Basic Service Manual Index 4...
Страница 643: ......