![GE LOGIQ S7 Expert Скачать руководство пользователя страница 70](http://html.mh-extra.com/html/ge/logiq-s7-expert/logiq-s7-expert_technical-publication_128038070.webp)
Site Preparations
3-18
Ultrasound System
–
Common Service Information
Direction 5444964-100 English
Rev. 5
EMI limitations
Ultrasound systems are susceptible to Electromagnetic
Interference (EMI) from radio frequencies, magnetic fields, and
transients in the air or wiring. They also generate EMI. The
Ultrasound system complies with limits as stated on the EMC
label. However, there is no guarantee that interference will not
occur in an installation.
Possible EMI sources should be identified before the Ultrasound
system is installed.
Electrical and electronic equipment may produce EMI
unintentionally as the result of a defect. Some of these sources
include:
• medical
lasers
• scanners
• cauterizing
guns
• computers
• monitors
• fans
• gel
warmers
• microwave
ovens
• light
dimmers
• mobile
phones
•
in-house wireless phones (DECT phones)
•
wireless computer keyboard and mouse
•
air conditioning system
•
High Frequency (HF) surgery equipment
• general
AC/DC
adapters
The presence of a broadcast station or broadcast van may also
cause interference.
See: ‘EMI prevention/abatement’ on
for EMI
prevention tips.
Содержание LOGIQ S7 Expert
Страница 2: ... 2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com ...
Страница 6: ...i 4 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 7: ...Ultrasound System Common Service Information i 5 Direction 5444964 100 English Rev 5 ...
Страница 8: ...i 6 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 9: ...Ultrasound System Common Service Information i 7 Direction 5444964 100 English Rev 5 ...
Страница 10: ...i 8 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 11: ...Ultrasound System Common Service Information i 9 Direction 5444964 100 English Rev 5 ...
Страница 12: ...i 10 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 13: ...Ultrasound System Common Service Information i 11 Direction 5444964 100 English Rev 5 ...
Страница 14: ...i 12 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 15: ...Ultrasound System Common Service Information i 13 Direction 5444964 100 English Rev 5 ...
Страница 20: ...i 18 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 24: ...i 22 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 52: ...Safety Information 2 24 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 78: ...Site Preparations 3 26 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5 ...
Страница 127: ... 2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com ...