8-76
Section 8-11 - Replacement of Monitor and LCD Arm Plastic Covers
8-11-6
LCD Arm Plastic Covers Remove Procedure
1) Unscrew 1 screw, than remove 3rd Arm Cover. Refer to the figure below.
2) Remove 2nd Axis Cover. Refer to the figure below.
3) Remove 2nd Arm Cover. Refer to the figure below.
Figure 8-64 Removing the 3rd Arm Cover
Figure 8-65 Removing 2nd Axis Cover
Figure 8-66 Removing 2nd Arm Cover