Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 65
8-2-26
DTRX64II Assy
8-2-26-1
Tools
•
Common philips screwdrivers
8-2-26-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-26-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
.
2.) Remove the EMI Cover R. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
.
3.) Unscrew 2 screws (1-2) and eject the PCB. Refer to the Figure 8-105.
4.) The DTRX64II Assys are settled in #1 and #2 slots in nest box.
5.) Pull the DTRX64II assy(s) out. Refer to the Figure 8-106.
Figure 8-105 Eject RFC Assy
Figure 8-106 DTRX64II Assy