Chapter 8 -
Page 8-85
8-2-38
ECG Cable Hook
8-2-38-1
Tools
•
Common Phillips screwdrivers
8-2-38-2
Removal procedure
1.) Unscrew 2 screws (1-2) to remove the ECG Cable Hook.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-38-3
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-126 Cable hook on the bottom of keyboard
Table 8-39 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-38. Equipment
passed all required tests and is ready for use. “
Cover Parts Function Validation
(1)
(2)