LOGIQ P3 S
ERVICE
M
ANUAL
3-10
Section 3-5 - Completing the Installation
Section 3-5
Completing the Installation
3-5-1
Power On / Boot Up
NOTE:
After turning off a system, wait at least twenty seconds before turning it on again. The system may not
be able to boot if power is recycled too quickly.
3-5-1-1
Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Ensure the Cable Clip slips securely over the shoulders on the molded plug.
3.) Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage.
Never use an adapter that would defeat the safety ground.
4.) Switch ON the Circuit Breaker at the rear of the system.
Figure 3-7 Scanner Power On
a. Circuit Breaker
b. Power Cable
When power is applied to the scanner and the rear Circuit Breaker is turned ON, power is distributed to
the Transformer Assembly and to the TMST.