GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
Chapter 4 Functional Checks
4-3
When power is applied to the scanner, power is distributed to the Cooling Unit, Control Panel, LCD,
Peripherals and the TMST.
4-3-1-2
Turn on the system
Press the
Power On/Off
switch at the front of the system once.
When the
Power On/Off
switch on the Control Panel is pressed once, the TMST starts and the software
code is distributed to initiate the scanner.
No status messages are displayed during this process.
4-3-2
Power Off/ Shutdown
NOTE:
After turning off a system, wait for at least twenty seconds before turning it on again. The system may
not be able to boot if power is recycled too quickly.
4-3-2-1
TMST Power Down
To power down the system:
1.) Press the
Power On/Off
switch at the front of the system once.
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select
key, select Shutdown.
Figure 4-2 Power On/Off Switch
Figure 4-3 System Exit Window
Power On/Off
switch
Содержание LOGIQ P3
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2 ...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii ...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork ...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting ...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software ...