GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
8-32
Section 8-4 - Mechanicals
Section 8-4Mechanicals
8-4-1
RH Side Cover (FRU No :5315115)
This is a description on how to remove and replace the Right Cover.
8-4-1-1
Tools
•
Common Torx screwdrivers
8-4-1-2
Needed Manpower
•
1 person, 15 minutes
8-4-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-1-4
Removal Procedure
1) Remove the screw cap of four places(1,2,3,4,) on RH-side cover of the system.
Refer
2) Remove the RH-side cover from the system. Refer
Figure 8-34 RH-side cover
3) Remove the RH-side cover four screws(1, 2, 3, 4) of the system. Refer
Screw Cap
RH-Cover