GE
PROPRIETARY TO GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-9
8-2-6
Handle P3 Console (FRU P/N: 5315120)
This is a description on how to remove and replace the Handle P3 console Assembly.
8-2-6-1
Tools
•
Common phillips screwdrivers; Allen Key Set.
8-2-6-2
Needed Manpower
•
1 person, 15 minutes
8-2-6-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-2-6-4
Removal procedure
1) Remove keyboard Top cover and Pull out the keyboard top assembly cover by the slot marked in the
2) Remove Gel holder and probe holder. Refer
) Using the allen key set remove
Handle P3 console by unfastening the two socket head fasteners
marked (1,2,3)
Figure 8-9 Handle P3 console
8-2-6-5
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-2-6-6
Functional Checkout Procedure
Functional Test
Debrief Script
Do Visual Inspection and ensure
no scratches observed and
fitment is done properly
Service Manual Direction 5308917-100, Section 8-2-5. Equipment passes all required
tests and is ready for use.
handle P3 console
Gel holder
Probe holder
Содержание LOGIQ P3 PRO
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software...