Facility needs
LOGIQ e – Basic Service Manual
2-13
5461614-100 English Rev. 6
Required facility needs
NOTE:
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
Sites with a mains power system with defined Neutral and Live
:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full size ground wire from
the distribution panel to the Ultrasound outlet
.
Sites with a mains power system without a defined Neutral
:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full size ground wire from the
distribution panel to the Ultrasound outlet
.
NOTE:
Please note that image artifacts can occur, if at any time within
the facility, the ground from the main facility's incoming power
source to the Ultrasound unit is only a conduit.
•
Dedicated single branch power outlet of adequate
amperage meeting all local and national codes which is
located less than 2.5 m (8 ft.) from the unit’s proposed
location
•
Door opening is at least 76 cm (30 in) wide
•
Proposed location for unit is at least 0.2m (0.67 ft.) from the
wall for cooling
•
Power outlet and place for any external peripheral are within
2 m (6.5 ft.) of each other with peripheral within 1 m of the
unit to connect cables.
•
Power outlets for other medical equipment and gel warmer
•
Power outlets for test equipment and modem within 1 m (3.2
ft.) of unit
•
Clean and protected space to store transducers (in their
cases or on a rack)
•
Material to safely clean probes (done with a plastic
container, never metal)
For the amperage requirements, see:
Содержание LOGIQ e
Страница 4: ...i 2 LOGIQ e Basic Service Manual 5461614 100 English Rev 6 Important precautions Translation policy...
Страница 5: ...LOGIQ e Basic Service Manual i 3 5461614 100 English Rev 6...
Страница 6: ...i 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 7: ...LOGIQ e Basic Service Manual i 5 5461614 100 English Rev 6...
Страница 8: ...i 6 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 9: ...LOGIQ e Basic Service Manual i 7 5461614 100 English Rev 6...
Страница 10: ...i 8 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 11: ...LOGIQ e Basic Service Manual i 9 5461614 100 English Rev 6...
Страница 12: ...i 10 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 24: ...i 22 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 60: ...Introduction 1 36 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 80: ...Site Preparations 2 20 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 134: ...System Setup 3 54 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 208: ...Components and Functions Theory 5 14 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 262: ...Replacement Procedures 8 26 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 328: ...Care and Maintenance 10 44 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 384: ...Docking Cart Setup 11 56 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 426: ...Index 4 LOGIQ e Basic Service Manual 5461614 100 English Rev 6...
Страница 427: ...GE...