10-3
10-3-1
How often should care & maintenance tasks be performed?
The Care & Maintenance Task Schedule (provided on page
) specifies how often your
LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e should be serviced and outlines items requiring special
attention.
NOTE:
It is the customer’s responsibility to ensure the LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e care &
maintenance is performed as scheduled in order to retain its high level of safety, dependability
and performance.
Your GE Service Representative has an in-depth knowlegde of your LOGIQ e/LOGIQ e Vet/LOGIQ i/
Vivid e ultrasound scanning system and can best provide competent, efficient service. Please contact
us for coverage information and/or price for service.
The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule
assumes that you use your LOGIQ e/LOGIQ e Vet/LOGIQ i/Vivid e for an average patient load (10-12
per day) and use it as a primary mobile unit which is transported between diagnostic facilities.
NOTE:
If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the maintenance frequencies.
NOTE:
May require specialized equipment to complete
NOTE:
PMs are not mandatory, the table above is for reference only.
Table 10-2 Customer Care Schedule
Service at Indicated Time
Daily
Weekly
Monthly
Per Facilities QA
Program
Notes
Clean Probes
•*
* or before each use
Inspect AC Mains Cable
•
Mobile Unit Check Weekly
Inspect Cables and Connectors
•
Clean Console
•
Clean LCD
•
Console Leakage Current Checks
See Note
Twice Annually
Peripheral Leakage Current Checks
See Note
Twice Annually
Surface Probe Leakage Current Checks
See Note
Twice Annually
Endocavity Probe Leakage
Current Checks
See Note
Quarterly
Annually
Measurement Accuracy Checks
See Note
Twice Annually