Optimizing Spectral Doppler
LOGIQ 7 Online Help
5
-
99
Direction 2392536-100 Rev. 1
PRF
Description
Adjusts the velocity scale to accommodate faster/slower blood
flow velocities. Velocity scale determines pulse repetition
frequency.
If the sample volume gate range exceeds single gate PRF
capability, the system automatically switches to high PRF mode.
Multiple gates appear, and HPRF is indicated on the display.
High PRF
High Pulse Repetition Frequency (HPRF) is a special operating
mode of PW Doppler. In HPRF mode, multiple energy pulses
are used. This allows higher velocities to be detected without
causing aliasing artifacts. HPRF mode is used when detected
velocities exceed the processing capabilities of the currently
selected PW Doppler scale or when the selected anatomical site
is too deep for the selected PW Doppler scale. The pulse
repetition frequency (PRF) is displayed to the left of the
spectrum in frames per second.
NOTE:
Ensure that only one gate overlays a blood vessel at a time.
Otherwise, signals from more than one flow area are
superimposed.
Adjusting
To raise/lower, adjust the
PRF
on the Touch Panel (PRF and
Wall Filter share the same control. Press the control to toggle
between PRF and Wall Filter). The display updates velocity
scale parameters after you adjust the velocity scale.
Values
Velocity Scale values vary by probe and application. In Triplex,
when you change the velocity scale in Color Flow, the Doppler
Mode velocity scale is also updated.
Velocity Scale values are returned to the factory or user preset
value when you change the following: Probe, Exam Category,
Exam Calcs, or New Patient.
Benefits
Blood flow information is not cut off due to the effect of aliasing.
Affect on other
controls
When you raise the velocity scale, the spectral waveform may
decrease in size; when you lower the velocity scale, the spectral
waveform may increase in size. Changes in the spectrum are
relative to changes in the velocity scale, that is, it sizes
accordingly. When you adjust the velocity scale, CINE memory
is cleared. Adjustments may affect sample volume size and
Doppler wall filter.
Bioeffects
Changing the velocity range may change the TI and/or MI.
Observe the output display for possible effects.
Содержание LOGIQ 7
Страница 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Страница 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 400: ...General Measurements and Calculations 7 94 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 699: ...LOGIQ 7 Basic User Manual 14 1 Direction 2392206 100 Rev 1 Chapter 14 ReportWriter Describes how to generate reports ...
Страница 741: ...LOGIQ 7 Basic User Manual 15 1 Direction 2392206 100 Rev 1 Chapter 15 Recording Images Describes how to record images ...
Страница 779: ...Presets LOGIQ 7 Basic User Manual 16 3 Direction 2392206 100 Rev 1 Overview continued Figure 16 1 Utility Touch Panel ...
Страница 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 983: ...GE Medical Systems ...
Страница 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...