GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8 - 183
8-11-1-1
System Checks
1.) Approx. two minutes later, the windows XP desktop appears. Then one minutes later, the following
screen appears. Clisck on
Yes
using right button of the trackball to restart the system
NOTE:
If the window above is not displayed, go to next step
2.) Drive Letter Checks:
a.) Double click on My Computer
b.) First check to see if you need to change the drive letters on the CDROM drive and the
Magneto Optical drives – they
SHOULD
be G: and H: respectively.If they are, go to next
step 3.) Driver Checks
Figure 8-212 System Setting Change
Figure 8-213 My Computer
Содержание LOGIQ 5 PRO
Страница 2: ...GE MEDICAL SYSTEMS DIRECTION 2407381 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL Page 1 2 ...
Страница 3: ......
Страница 7: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL ii iii ...
Страница 50: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL 2 12 Section 2 3 Facility Needs ...
Страница 462: ...GE MEDICAL SYSTEMS DIRECTION 2380207 REVISION 7 LOGIQ 5 PRO SERVICE MANUAL A 4 Index INDEX ...