GE M
EDICAL
S
YSTEMS
D
IRECTION
2380207
, R
EVISION
7
LOGIQ™ 5 PRO S
ERVICE
M
ANUAL
8 - 158
Section 8-9 - Power Block
8-9-3
JPC Assy (FRU No. 702)
Purpose: This is a description on how to remove and replace the JPC Assy.
8-9-3-1
Tools
•
Common pillips screwdrivers
•
Allen/Unbraco wrench
8-9-3-2
Needed Manpower
•
1 person, 15 m travel
8-9-3-3
Preparations
•
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
8-9-3-4
Removal Procedure
1.) Unscrew nine (9) hexagon screws (1-9) to remove the AC Power Top cover.
2.) Unscrew five (5) screws (10-14).
CAUTION
Do not wear the ESD wrist strap when you remove a part of power supply unit. Turn OFF power
and unplug the power cord before removing a part of power supply unit. However be sure to turn
off power and wear the strap before you remove a circuit boards.
Figure 8-176 JPC Assy
JPC Assy