Facility needs
Ultrasound System
–
Common Service Information
3-23
Direction 5444964-100 English
Rev. 5
Required facility needs
NOTE:
GE requires a dedicated power and ground for the proper
operation of its Ultrasound equipment. This dedicated power
shall originate at the last distribution panel before the Ultrasound
system.
The Ultrasound system will function on voltages from 100-240
Volts and 50 or 60 Hz
.
However, if using 220-volt power in North
America, then a center tapped power source is required
.
Sites with a mains power system with defined Neutral and Live
:
The dedicated line shall consist of one phase, a neutral (not
shared with any other circuit), and a full-size ground wire from
the distribution panel to the Ultrasound outlet
.
Sites with a mains power system without a defined Neutral
:
The dedicated line shall consist of one phase (two lines), not
shared with any other circuit, and a full-size ground wire from
the distribution panel to the Ultrasound outlet
.
Figure 3-1.
Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
Scale
:
Each square equals one square
foot (app. 31 x 31 cm)
1. Door – at least 762 mm (30
inches)
2. Film
Viewer
3. Counter Top, Sink with hot
and cold water and Supplies
Storage
4. Linen
Supply
5. Probes/Supplies
6. Examination Table – 1930 x
610 mm (76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound
system
10. External Peripherals
11. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
12. Network Interface
13. 457 mm (18 inches) distance
of Ultrasound system from
wall or objects
14. GE Cabinet for Software and
Manuals
4
1
3
11
6
13
8
10
5
7
9
12
14
2
Содержание LOGIC C2 Series
Страница 2: ...2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com...
Страница 6: ...i 4 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 7: ...Ultrasound System Common Service Information i 5 Direction 5444964 100 English Rev 5...
Страница 8: ...i 6 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 9: ...Ultrasound System Common Service Information i 7 Direction 5444964 100 English Rev 5...
Страница 10: ...i 8 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 11: ...Ultrasound System Common Service Information i 9 Direction 5444964 100 English Rev 5...
Страница 12: ...i 10 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 13: ...Ultrasound System Common Service Information i 11 Direction 5444964 100 English Rev 5...
Страница 14: ...i 12 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 15: ...Ultrasound System Common Service Information i 13 Direction 5444964 100 English Rev 5...
Страница 20: ...i 18 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 24: ...i 22 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 52: ...Safety Information 2 24 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 78: ...Site Preparations 3 26 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Страница 127: ...2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com...