2
SAFETY
2.1
Symbols
The meaning of symbols used on the Kelman
™
DGA 900 Plus:
Caution. Refer to the Installation Manual / Operator Guide to prevent death,
injury, equipment damage or loss of data.
Electrical Hazard. Risk of electric shock.
Primary Protective Earth connection.
Hot surfaces may be present.
The meaning of symbols used in this manual:
WARNING
A procedure, practice, or condition could cause death or serious injury
and/or significant equipment damage.
A procedure, practice, or condition could cause minor injury, equipment
damage or loss of data.
Electrical Hazard: Risk of electric shock.
2.2
Safety Statements
The following safety statements must be observed:
The customer and installer are responsible for ensuring that all local
regulations and site policies are complied with concerning safe working
practices.
The minimum ambient temperature for installation and service activities is
−10 °C.
.
If the equipment is installed or used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.
If working at height, third parties must have received appropriate training
for working at height prior to work commencing. This includes, but is not
limited to, ‘Working at height’ and ‘Using Mobile Elevated Working
Platforms’ training.