Valves IV-908, PV-908

Instructions

GE Healthcare

Содержание IV-908

Страница 1: ...Valves IV 908 PV 908 Instructions GE Healthcare...

Страница 2: ...s A copy of the corresponding Declaration of Conformity is available on request The CE symbol and corresponding declaration of conformity is valid for the instrument when it is used as a stand alone u...

Страница 3: ...the low pressure side before the pump and PV 908 on the high pressure side after the pump The valve may be used to switch between sample and buffer solutions Two valves may be used to switch between...

Страница 4: ...4 Instructions 56 8101 02 Edition AE 1 Introduction...

Страница 5: ...ould correspond to that used in UNICORN Note All valves must have different numbers 3 Mount the valve vertically in the way it is shown in the user documentation of your KTAdesign chromatography syste...

Страница 6: ...6 Instructions 56 8101 02 Edition AE 2 Installation...

Страница 7: ...Edition AE Operation 3 3 Operation The valves are controlled from UNICORN Storage Overnight The valves can be left filled with a buffer Overnight and Long time storage Flush the valves with water and...

Страница 8: ...3 Operation 8 Instructions 56 8101 02 Edition AE...

Страница 9: ...t is available see ordering information in Reference Information 1 Ensure that the valve is disconnected from the pump and the tubing are disconnected 2 Remove the 4 screws on the front using the supp...

Страница 10: ...l indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately Please contact an authorized representative of the m...

Страница 11: ...g to wrong position The valve parts may have been incorrectly assembled after replacement 1 Check that the distribution plate marking i o is horizontal External leakage 1 Check the tubing connections...

Страница 12: ...mbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately Please contact an authorized representative of th...

Страница 13: ...ed to one of the peripheral ports 1 8 allowing a clear liquid path Valve switching is controlled from UNICORN by reading the actual position of the channel plate The material used in the switching par...

Страница 14: ...hy of biomolecules except 100 ethylacetate 100 hexane and 100 tetrahydrofuran THF Power requirement 32 V DC 10 from the system pump Power consumption Up to 9 W UniNet 2 address 0 9 Inlet and outlet tu...

Страница 15: ...er Plates 0 9 1 18 1109 09 Mounting bracket 1 18 1109 11 Teflon TM tubing i d 1 8 o d 3 16 3 m 18 1112 47 Tubing connector for 3 16 o d tubing 10 18 1112 49 Ferrule for 3 16 o d tubing 10 18 1112 48 S...

Страница 16: ...way NJ 08855 1327 USA GE Healthcare Bio Sciences KK Sanken Bldg 3 25 1 Hyakunincho Shinjuku ku Tokyo 169 0073 Japan Asia Pacific Tel 85 65 62751830 Fax 85 65 62751829 Australasia Tel 61 2 8820 8299 Fa...