Base plate assembly (mandatory) 2285048 – Refer to
•
If there is no integrated baseplate plan (2359194) delivered with S18101SG, please refer to
drawings 2359194 (Angio configuration
) or 2359194 (Cardio configuration
). Please order locally the integrated base plate based on this plan.
•
Assembly and separation (Select kit) 2285050 – Refer to tables,
○
Above grade, Floor anchors without insert – 2285051.
○
Through bolts without insert – 2285646.
○
On grade, Floor anchors – 2286398.
○
Above grade, Floor anchors and insert – 2285052.
○
Through bolts and insert – 2285053.
○
Floor plate / Base plate Assembly – 2285632.
•
Templates (select kit) – 2285054 – Refer to
Angio/Cardio/Neuro templates – 2285056 (containing 2185979, 2127792 and 2360133
templates).
1.1 2285048 - Base Plate Assembly
Table 2-1:
Item
Name
Part #
Description
Quan.
Notes
Baseplate
2285059
12 Hole Floor Baseplate
1 pc
Lift Plate
2290939
Lifting Straps
2 pc
Hex Head
Screw
5166535
Hex Head Screw 1/2”-13UNC L1
1/2” 12.9 Black
2 pc
Grease
2295599
Lithium grease lubricant 100g
2 pc
Doc
2230112-1-100
Vascular Gantry Baseplate and Ta‐
ble
1 pc
Doc
2229297-100
LCA/LCV+/LC+ System Baseplate
and Omega IV/V/EP Table Floor
Preparation
1 pc
Doc
2290880-2-100
Innova Pre-Installation Kit Install
Procedure
1 pc
NOTE:
The 12 Gantry baseplate mounting screws are delivered with the system's Gantry. These
screws are:
•
cap screws 5166774; Screw Socket Head Cap 1/2”-20 UNF L1 1/2” 12.9 Black (used
when installing new system on old baseplate (US threads))
•
cap screws 2300939; Hex Head Cap screw; M12 40/40 Class 12.9 Black ; used to
attach L-brackets on table shipping pallet to positioner dolly for table positioning
(used when the new baseplate is metric threads)
Single Plane and Biplane Innova Systems Pre–Installation Kit Installation Procedures