11 Injectors
The injector is provided with an additional ground cable (P/N #2135737) to meet equipotentiality
requirement at patient vicinity.
Injector emergency power off:
•
if no emergency power off is required for the injector, the injector can be powered from a
wall outlet socket.
•
if emergency power off circuit must include the injector, ensure the power for the injector is
via the PDB.
11.1 Remote Injector (rack mount)
A power cord is supplied with the injector.
110 VAC or 230 VAC are supplied from the main disconnect room device. The injector must be
connected to this supply.
11.2 Pedestal Injector
A power cord is supplied with the injector.
In all cases (110 VAC or 230 VAC), it will be connected to a wall outlet near the operator
location, fed by the main disconnect room device.
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2
Chapter 5 Electrical Requirements
201