10 Advantage Windows
A power cord is supplied with Advantage Windows. It is connected to outlet fed by the main
disconnect room device.
Verify that the ground between AW, AW spliter and monitors ground are equipotential (0.1 ohm
equipotentiallity between room ground point (busbar) and wall outlets).
NOTICE
To power the AW splitter, use the appropiate cable out of lot P/N 2385173.
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2