7 Door Interlocks
CAUTION
IEC 60601–2–43 requires not to install door interlocks. It is the responsibility of
the field service to check that this requirement is not in contradiction with local
regulation. In case of conflict, follow local regulation. No other measures
employed for radiation protection should cause the interruption of irradiation and
any other disturbance of a procedure in progress.
Illustration 5-23: C2 cabinet bulkhead RM DOOR connecting
Innova
TM
IGS 620, Innova
TM
IGS 630 Preinstallation Manual
Direction 5694389-1-1EN, Revision 2
Chapter 5 Electrical Requirements
197