3 Heat Output
3.1 Equipment Heat Output tables
. To obtain heat output information for components not specified in
,
refer to the appropriate component Pre–Installation Manual listed in
Table 4-4:
HEAT OUTPUT
Stand by
Moderate Use (4)
Typical Use (4)
Maximum Use (4)
Room
Core System
kW
BTU/hr
kW
BTU/hr
kW
BTU/hr
kW
BTU/hr
Exam
Room
Innova LC positioner and
table
0.41
1394
0.55
1858
0.89
3020
1.62
5517
Ctrl
Room
DL user area with 1 TFT
monitor
0.16
546
0.16
546
0.16
546
0.16
546
2 B&W flat monitors
0.17
573
0.17
573
0.17
573
0.17
573
Tech
Room
C1 Cabinet
0.41
1398
0.69
2366
0.99
3389
1.29
4412
C2 Cabinet
0.29
989
0.33
1125
0.54
1828
0.87
2966
Coolix X-Ray tube chiller
(1) (2)
2.53
8619
4.49
15309
5.49
18725
6.93
23625
Detector conditioner
0.21
709
0.21
709
0.21
709
0.21
709
Main disconnect panel -
PDB
0.4
1534
0.45
1534
0.45
1534
0.45
1534
Total for core system
4.6
15762
7.0
24020
8.9
30324
11.7
39881
Room
Options (3 & 5)
Stand by
Moderate Use (4)
Typical Use (4)
Maximum Use (4)
Exam
Room
3 in room B&W TFT mon‐
itors
0.25
859
Same values as
Stand by
Same values as
Stand by
Same values as
Stand by
In room AW TFT monitor 0.12
409
Typical injector
0.09
320
LD Monitor
0.5
1706
Ctrl
Room
AW work station
0.35
1201
2 AW TFT monitors
0.24
818
Printer
0.31
1054
LD Monitor (6)
0.5
1706
Tech
Room
LD Cabinet
0.15
512
3kVA Cabinet UPS -
model 9130 (7)
0.37
1257
Typical configuration without fluoro
UPS
5.9
20424
8.4
28682
10.3
34985
13.1
44543
NOTE:
(1) Air flow requirements 1200 m
3
/h (706 CFM)
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
190