During the X-Ray operation, the Vascular gantry movements also limit the Vivid E9
positions, and a dedicated back out path has to be cleared for Vivid E9 park position.
2.3.4 Peripheral Equipment
Consult hospital personnel regarding additional space requirements for the following types of
hospital equipment:
•
Storage cabinets.
•
Sinks.
•
Oxygen stations.
•
IV apparatus.
•
Injectors.
•
Heart monitoring equipment.
•
Crash cart.
2.3.5 Emergency Stop
Protect the Emergency Stop from accidental actuation.
2.3.6 Patient Environment Equipment
WARNING
OPTIONAL LARGE DISPLAY SECONDARY MONITOR ON WALL-MOUNTED
SUSPENSION MUST BE INSTALLED OUTSIDE OF THE PATIENT VICINITY
The components that may be installed within patient vicinity need to be medical equipment
(”patient vicinity is defined in the standardization as a space within the room 1.83 m (70.7 in)
beyond the perimeter of the examination table and extending vertically 2.29 m (90.2 in) above
the floor.”). For the Innova System, the equipment are:
•
Table
•
C-arm
•
Monitors
•
Injector
•
Rad-Shield
•
Intra Vascular Ultrasound System (IVUS)
○
IVUS Control Console.
○
IVUS Joystick.
○
IVUS Touch Pad
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 2 Equipment Requirements
125