HygroTrace Moisture Transmitter
5
September 2007
Standard Wiring Connections (cont.)
2.
Using the flying leads at the other end of the transmitter cable,
connect the transmitter to the power supply and data
acquisition system (DAS) as shown in Figure 4 on page 6.
Refer to Table 1 below for a description of the leads in the
factory-supplied cable.
3.
Trim any unused leads back to the outer cable jacket in order
to remove the bare tinned wire and prevent accidental short
circuits.
The
HygroTrace
is now ready for operation.
Note:
To remove the probe cable, pull back the spring-loaded
sleeve and pull the cable out of the connector.
Table 1: Cable Lead Connections - No PC
Lead
Connection Description
Red
(+) 24VDC 1 AMP
Black
(–) 24VDC 1 AMP
Orange
(+) 4-20mA Output
Blue
(–) 4-20mA Output
Cable Braid
System Ground