GE Protimeter Hygromaster
Page 17
7.
Enter the required logging parameters as follows:
Note:
Be sure to enter the Logging Interval, Wait Before First Log, Number of Logs and Job
No in the appropriate boxes.
a.
Press
“Set Interval”
button to send to the instrument.
b.
Confirm that the logging parameters are set correctly by pressing the
“Check Interval”
button, and ensure that the settings you require have appeared in the appropriate
parameter boxes.
c.
Finally, check that the ‘logging’ icon
is lit on the instrument display.
Note:
It does not matter whether the above is set from the PC or in
“Logging Setup”
mode, as
described in Logging Set-up Function (Table 6).
8.
Remove the RS232 serial cable from the instrument port and place it in the environment that
is to be monitored. Ensure that a Hygrostick or Quikstick sensor is plugged into the top of
the instrument.
9.
Press and hold
until the display shows
OFF
.