3.2.6
Fluoroscopy
3.2.7
eXplore Locus SP Electrical Block Diagram
The following provides an electrical block diagram that may be used when troubleshooting the
eXplore Locus SP.
Fluoroscopy
Problem
Possible Cause
Suggested Recovery/Solution
No Fluoro
Are Fluoro Window/Level slide bars
correctly set?
If not, adjust slide bars or check filter.
Are X-Rays on?
If not, turn on X-Ray Source.
If X-Ray Source can not be turned on,
check whether eXplore Locus SP has mis-
configured jumpers.
If jumpers are not mis-configured, check
X-Ray tube operation.
Is shutter working?
Open the lid and observe if there is a
flashing red LED. If not, check for
’ShutterServer.exe in the process list.
If ShutterServer.exe does not exist, restart
shutter (’net stop shtrsvr’, ’net start
shtsvr’).
If shutter is still not working, restart the
scanner and restart the computer.
Is computer hung?
Is computer is hung, reboot computer.
Possible hardware connection
problem
Check frame-grabber cable.
Possible protocol mismatch to
hardware
Check for mismatch (i.e. edge trigger
protocols vs. internal trigger protocols).
Error message ’cannot connect to
camera’:
•
Check whether the scanner is
powered.
•
Check power supply (fuses).
•
Check serial connection between PC
and detector.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...