Section 7.3 Stage Replacement Calibration
The following describes the calibration procedures performed each time the stage is replaced. The
procedures below use the following applications:
•
System Warm-Up
(C:\Program Files\GE Medical Systems\eXplore Utilities\WarmupApp.exe)
•
Calibration Tools
(C:\Program Files\GE Medical Systems\eXplore Utilities\Calibration.exe)
•
X-Ray and Motor Controller
. The
X-Ray and Motor Controller
(C:\Program Files\GE
Medical Systems\eXplore Utilities\X_Ray_Controller.exe)
7.3.1
Mechanically Align the Stage
1.
Open
Windows Explorer
and navigate to D:/. Copy the
calib
folder and rename it
original_calib
.
2.
Place the 9" Square tool on the stage with one edge flat against the detector.