Section 5-5 - Common Service Interface
5-5-6
Diagnostics
Detailed
Diagnostic
information is found in
Section "Diagnostics/Troubleshooting" on page 7-1
5-5-6-1
Diagnostics Execution
Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level
diagnostic selection based on the user’s level and login access permissions. Remote access will require
disruptive diagnostic permissions to run Acquisition diagnostics.
5-5-6-2
Diagnostic Reports
Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for
future viewing of the diagnostic logs.
5-5-6-3
Proactive Diagnostics
A system of self-monitoring is largely supported with the integration of
iLinq
. The scheduler, executive,
user interfaces and some of the proactive diagnostic functions are provided by
iLinq
. Other tasks should
be provided by the product team.
Figure 5-23 GE Service Diagnostic Page