8-100
Section 8-6 - Mechanical Parts- Replacement Procedures
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-6-8
MPB Blower (Fan) Replacement Procedure
8-6-8-1
Tools
Appropriate Phillips screwdriver
8-6-8-2
Time Required
10 min
8-6-8-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-8-4
MPB Blower (Fan) Removal Procedure
1.) Remove all accessories.
2.) Remove the following covers: Lower Front eTower, Left side eTower, and Right Side eTower,
3) Remove MPB Front Metal Door.
4.) Disconnect MPB to BIB Control Cable from the fan cable connector.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
RS Probe Cover Removal Procedure
•
Lower Front eTower Cover Removal Procedure
•
Left Side eTower Cover Removal Procedure
•
Right Side eTower Cover Removal Procedure
•
MPB Front Metal Door Removal Procedure
Figure 8-85 Disconnecting Fan Cable Connector
MPB to BIB
Control Cable
Содержание H45051VN
Страница 477: ......