Chapter 8 - Replacement Procedures
8-83
D
IRECTION
5771498-100, R
EVISION
6
V
ENUE
™ S
ERVICE
M
ANUAL
P R E L I M I N A R Y
8-6-3
Halo Handle Replacement Procedure
8-6-3-1
Tools
LOCTITE® 243™
Allen key 3 mm.
8-6-3-2
Time Required
30 min
8-6-3-3
Preparations
1.) Shut down the Venue™ ultrasound unit, as described in
.
2.) Make sure the On/Off power switch is set to Off.
3.) Make sure the system is standing securely on a level surface, with the wheels in the locked position.
8-6-3-4
Halo Handle Removal Procedure
1.) Remove all accessories.
2.) Remove cockpit (monitor).
3.) Remove the arm.
4.) Remove the Riser Thermal Cover.
5.) Remove the Riser Cover.
FRU Part #
Refer to
•
Accessories - Replacement Procedures
•
Cockpit (Monitor) Removal Procedure
•
•
Riser Thermal Cover Removal Procedure
•