Appendix B: Specifications
B-2
Service Manual
Installation
Non-Permanent Installation of Equipment
System Performance
Expected Service Life
Approximately 7 years.
The device is designed to last at least 7 years in normal use when operated,
maintained, and serviced in accordance with the instructions provided in the
provided manuals. With proper maintenance and repairs, the service life can be
extended as long as service parts are available.
Patient Temperature Measurement
Accuracy*
± 0.3°C @ 30°C to 42°C
Accuracy of patient temperature
measurement system within
range of temperature
measurement.
Temperature Control Accuracy*
± 1.0°C
Control temperature vs average
incubator temperature with level
bed in Manual Mode.
Variability
± 0.5°C
Incubator temperature vs
average incubator temperature.
Warm-up Time
< 50 min.
Time to reach 39°C control temp
from cold start in 25°C 50% RH
room ambient.
Unit with low noise fan: < 30 min.
Measured in accordance with
IEC 60601-2-19.
Air Velocity
≤
10 cm/s
In Whisper Quiet™ mode velocity
measured 10 cm above the
center of the mattress, closed
bed.
CO
2
Level
0.3%
Maximum CO
2
level measured per
IEC 60601-2-19.
Sound Level (within patient
compartment)
Unit with low noise fan:
≤
41 dBA
Unit without low noise fan: < 50 dBA
In Whisper Quiet™ mode closed
bed sound level measured 10 cm
above the center of the mattress.
Sound Level for Alarms
63 dBA (High Priority)
60 dBA (Medium Priority)
Sound pressure level is measured
in a free field over a reflecting
plane in accordance with
IEC 60601-1-8.
USB Port Output
0.5 A, 5 VDC
Note:
The above items noted with an asterisk (*) indicate essential performance.
Patient Contact Surfaces
Mattress
Skin Temperature Probe
Содержание Giraffe OmniBed Carestation CS1
Страница 4: ...RH 2 Service Manual ...
Страница 24: ...Service Language Disclaimer viii Service Manual ...
Страница 40: ...Safety and Regulatory Information User Responsibility xxiv Service Manual ...
Страница 44: ...Chapter 1 Functional Description Control Board 1 2 Service Manual FIGURE 1 1 Block Diagram ...
Страница 74: ...Chapter 3 Calibration and Maintenance Line Voltage Calibration 3 12 Service Manual ...
Страница 82: ...Chapter 4 Understanding the Service Interface Upgrade 4 8 Service Manual ...
Страница 110: ...Chapter 5 Troubleshooting Diagnostic Readings 5 28 Service Manual ...
Страница 211: ...Chapter 7 Repair Procedures Display Harness Connections Service Manual 7 75 ...
Страница 212: ...Chapter 7 Repair Procedures Display Harness Connections 7 76 Service Manual ...
Страница 272: ...Chapter 8 Illustrated Parts Options 8 60 Service Manual FIGURE 8 36 Silo Support Assembly 6600 0853 800 ...
Страница 290: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 78 Service Manual FIGURE 8 45 Wiring Diagram Control Panel ...
Страница 291: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 79 FIGURE 8 46 Wiring Diagram Incubator and Warmer Heater ...
Страница 292: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 80 Service Manual FIGURE 8 47 Wiring Diagram Servo Humidifier ...
Страница 293: ...Chapter 8 Illustrated Parts Wiring Diagrams Service Manual 8 81 FIGURE 8 48 Wiring Diagram Servo Control Oxygen ...
Страница 294: ...Chapter 8 Illustrated Parts Wiring Diagrams 8 82 Service Manual ...
Страница 296: ...Appendix A Compartment and Skin Probe Characteristics A 2 Service Manual ...
Страница 304: ...Appendix C RS 232 Serial Data Nurse Call C 4 Service Manual ...