11.3 Light Pipe / O-ring / Ground Clip Replacement
1. Make sure the system power cord is unplugged from the power outlet.
2. Use the provided hex key tool to loosen the two screws on the mounting bracket in the rear side of the
system and detach the system from the dovetail rail of the bed (Refer to Figure 11-3).
3. Lay the system on its front side on a flat surface. Use the provided hex key tool to remove the mounting
screw that attaches the light pipe ground clip to the device and detach the light pipe (Refer to Figure 11-
4. To remove the O-rings, rotate the light pipe while pulling out the O-rings.
5. To remove the ground clip, hold the light pipe end close to the clip firmly and then use a pair of long nose
or pliers to pull out the clip.
6. Reverse steps to re-install.
CAUTION:
The LED light output is susceptible to dust and dirt. When the light pipe is detached from the
system, take care that dust or dirt does not get into the light pipe port on the light box. When
storing system with open light pipe port for an extended period, cover the light pipe port.
Figure 11-3
Figure 11-4
65
Chapter 11: Replacement Procedures
Содержание Giraffe Blue Spot PT Lite
Страница 1: ...Giraffe Blue Spot PT Lite Operation Maintenance and Service Manual GE Healthcare ...
Страница 7: ... 7 Part I Operation and Maintenance ...
Страница 8: ...8 Part I Operation and Maintenance This page is intentionally left blank ...
Страница 16: ...This page is intentionally left blank 16 Chapter 1 Product Overview ...
Страница 22: ...This page is intentionally left blank 22 Chapter 2 Product Setup and Operation ...
Страница 26: ...This page intentionally left blank 26 Chapter 3 Operator s Maintenance ...
Страница 32: ...This page intentionally left blank 32 ...
Страница 33: ... 33 Part II Service ...
Страница 34: ...34 Part II Service This page is intentionally left blank ...
Страница 44: ...44 Important Service Safety Information This page is intentionally left blank ...
Страница 50: ...This page is intentionally left blank 50 Chapter 7 Installation ...
Страница 55: ...Chapter 9 Calibration No calibration is required for the Giraffe Blue Spot PT Lite 55 ...
Страница 56: ...This page is intentionally left blank 56 Chapter 9 Calibration ...
Страница 72: ...This page is intentionally left blank 72 Chapter 11 Replacement Procedures ...
Страница 88: ...12 5 Wiring Diagram 88 Chapter 12 Service Parts ...
Страница 94: ...This page intentionally left blank 94 ...
Страница 95: ......