Warning and Precautions 

2102322-111 REV 1

 

 Page 5 

Warning and Precautions 

WARNING 

ELECTROSURGERY 

 

There is a risk of burns and injury to the patient. If an electro surgery device is 

used, disconnect the ECG leadwires from the patient. 

LEADWIRES 

 

ECG Leadwires present a possible strangulation hazard. To avoid possible strangulation, 

route all ECG Leadwires away from patient's throat. 

CONDUCTIVITY 

 

Electric shock or device malfunction may occur if electrodes contact conductive 

materials. Keep the conductive parts of lead electrodes and associated parts away from other 

conductive parts, including earth. Also make sure that no contact to other conductive parts is possible

if the electrodes loosen during recording. 

GENERAL DANGER TO THE PATIENT 

 

Instructions listed in this manual in no way supersede 

established medical practices concerning patient care. Perform the established medical practices 

under all circumstances. 

EXPLOSION HAZARD

Do not use in the presence of a flammable anesthetic mixture with air or oxygen 

or nitrous oxide

.

DEFIBRILLATION - Device is defibrillation protected when the original Norav Medical patient cable is 

used. However, as a safety precaution when possible, remove the electrodes before defibrillation. 

GENERAL DANGER TO THE PATIENT - The device is not designed for direct cardiac application. 

INFECTION RISK 

 

Reuse of disposable parts that come into contact with patients pose a risk of 

infecting patients. Do not reuse disposable parts that have had direct contact with the patient, such as 

ECG electrodes. 

INTERPRETATION HAZARD 

Computerized interpretation is only significant when used in conjunction

with clinical findings. A qualified physician must over read all computer-generated tracings. 

MAGNETIC AND ELECTRICAL INTERFERENCE 

Magnetic and electrical fields are capable of interfering 

with the proper performance of the device. For this reason, make sure that all external devices 

operated in the vicinity of the device comply with the relevant EMC requirements. X-ray equipment or 

MRI devices are possible sources of interference as they may emit higher levels of electromagnetic 

radiation. Ensure that there are no X-ray arrays, MRI or high-performance transformers are near the 

device. 

OPERATOR 

Medical technical equipment such as this system must only be used by qualified and 

trained personnel. Operate the unit only at clinics and hospitals. Do not use at home. 

OPERATION WITH OTHER DEVICES 

Other devices which are part of the system must meet the requirements of the Standard for 

Information Technology Equipment (IEC/EN 60950-1) and the Standard for Electrical Medical

Devices (IEC/EN 60601-1). 

The personal computer should be approved to the appropriate safety standard for non-

medical electrical equipment (IEC/EN 60950-1, or its national variants). 

Accessory equipment connected to the analogue and digital interfaces must be certified 

according to the respective IEC/EN standards (e.g. IEC/EN 60950-1 for data processing 

equipment and IEC/EN 60601-1 for medical equipment). Furthermore, all configurations shall 

comply with the valid version of the standard IEC/EN 60601-1. Therefore anybody, who 

connects additional equipment to the signal input or output connector to configure a 

medical system, must make sure that it complies with the standard. 

When using GEH-ECG 1200 in combination with any other equipment, refer to a qualified 

service technician for correct handling.

MIXING UP RECORDINGS 

To ensure that a recording is not assigned to the wrong patient DO NOT use 

more than one identical marked RF ID pair of 1200W/1200WR at the same wireless environment. 

Caution 

DAMAGE TO THE DEVICE THROUGH BATTERY LEAKAGE 

 Batteries may leak if left in an unused device 

for prolonged periods. If you intend to store the device for longer than one week, remove the battery 

from it. 

LEADWIRE DAMAGE 

 Bending or wrapping the ECG Leadwire can damage it. When attaching and 

affixing the ECG leadwire, make sure not to bend it excessively. Avoid coiling the ECG leadwire around 

the device, as this can damage the leadwire. 

DAMAGE TO THE DEVICE 

 You may only open the battery compartment of the patient unit. Do not use

force when handling the recorder. 

SAFETY ONLY WITH APPROVED ACCESSORIES 

 Safe and reliable operation of the device is only possible 

when using the supplied and approved accessories. 

DIFFICULTIES FINDING CAUSES FOR MALFUNCTIONS 

 To find and repair a malfunction, both device and 

ECG leadwires are needed. Remember to include the ECG leadwire when returning the device for 

service or repair. (Avoid wrapping the ECG leadwires around the device, as this can damage the 

leadwires.) Always use the same ECG leadwires sets with a device. If an institution has several devices 

and ECG leadwires sets, try to ensure that each device is matched with a specific ECG leadwires set. In 

this way, leadwire or recorder failures can be isolated and eliminated faster. In the event of apparent 

changes in the performance of the device, discontinue use immediately. Do not resume use until the 

device is approved by the manufacturer or by a representative of the manufacturer. 

Содержание GEH-ECG 1200

Страница 1: ...322 111 REV 1 NOTE The information in this manual only applies to the product GEH ECG 1200 Due to continuing product innovation specifications in this manual are subject to change without notice 2019 General Electric Company All rights reserved ...

Страница 2: ...e program other than the original design or intended use thereof is not advised and considered a misuse of the software product Norav Limited Warranty Norav products are warranted to be free from manufacturing and material defects for a period of one 1 year from the date of shipment from Norav or the dealer to the original purchaser Excluded from this warranty are expendable supply items including...

Страница 3: ...ols and Indicators 6 Battery Assembly 7 Belt Strap Assembly 8 Electrode Application 8 Patient Leadwires Assembly 9 Maintenance and Cleaning 10 Storage 10 Service 11 Calibration 11 Troubleshooting 11 ECG Cables and Accessories 12 Technical Specifications 12 Electromagnetic Emissions and Immunity Information 13 ...

Страница 4: ...e device or signify an important step or procedure that requires special attention Device Labels Symbols Definition Defibrillation proof type CF Applied part Caution Refer to operation manual NOTE On ME EQUIPMENT Follow instructions for use 2797 CE marking Manufacturer Name and Address Device Serial Number Device Reference Number Date of manufacture Contains MIC certified RF module FCC ID Contains...

Страница 5: ...hich are part of the system must meet the requirements of the Standard for Information Technology Equipment IEC EN 60950 1 and the Standard for Electrical Medical Devices IEC EN 60601 1 The personal computer should be approved to the appropriate safety standard for non medical electrical equipment IEC EN 60950 1 or its national variants Accessory equipment connected to the analogue and digital int...

Страница 6: ...nd This device must accept any interference received including interference that may cause undesired operation The manufacturer is not responsible for any Radio or TV interference caused by unauthorized modifications to this equipment Such modifications could void the user s authority to operate the equipment NOTE NO OPTIMUM WIRELESS PERFOMANCE Note that the complete data transfer cannot be guaran...

Страница 7: ...ore than two weeks when the recorder is not in use If you use rechargeable batteries the battery recharger should be kept out ofthe patient environment and hook up area Dispose of used batteries carefully using environmentally friendly methods wherever possible following the state s recycling laws or your facility srecycling policy Use only the recommended battery type as instructed in the technic...

Страница 8: ...suitable for use As ECG electrodes from different manufacturers have different electrical properties the choice of ECG electrodes can considerably affect the measurement results and quality Ensure that only high quality electrodes are used Wet gel electrodes are recommended Always refer to the ANSI AAMI EC12 2000 Standard for safety performance and labeling requirements for the disposable electrod...

Страница 9: ...eat height or wear and tear chafed patches on the cable Do not use the instrument or the leadwires if you detect cracks melted areas or any other signs of damage to the leadwires or housing Caution Verify that dates on applicable accessories have not expired ECG electrodes can cause skin irritation Examine the skin for signs of irritation or inflammation and avoid placement of the electrode in tho...

Страница 10: ... measures as the rejected items Use the recorder after all the items become accepted Unit check shall be performed by each medical institution or by Norav personnel representative agent or an authorized third party For more details do not hesitate to consult your dealer or Norav Medical personnel Details of the check Check Method Criteria Operation manual Check that the operation manual is kept in...

Страница 11: ...te before packing unless you suspect they are associated with the problem Wherever possible use the original shipping carton and packing materials Include a packing list and the Norav Return Material Authorization RMA number It is recommended that all returned goods be insured Claims for loss or damage to the product must be initiated by the sender Calibration The device does not need any calibrat...

Страница 12: ...500 Lead OFF Detection Yes Communication Digital RF 2400 2483 MHz 0 4mW conform to FCC Part 15 249 Power Battery Size AA x 2 Alkaline or NiMH Operation time Up to 40 hours with Alkaline Batteries Mechanical Size 140 x 95 x 50 mm Weight 350g Environmental Operating Temp Range 10 to 40 C Storage Temp Range 20 to 60 C Relative Humidity 10 to 95 non condensing 1200WR Receiver Auxiliary output ECG Out ...

Страница 13: ...N A Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec N A Mains power quality should be that of a typical commercial or hospital environment Po...

Страница 14: ...ween portable and mobile RF communications equipment and NR recorder This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled Users of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment transmitters and the device as recommended below according t...

Страница 15: ...Electromagnetic Emissions and Immunity Information 2102322 111 REV 1 Page 15 ...

Страница 16: ...www gehealthcare com ...

Отзывы: