Chapter 8 - Replacement Procedures
8-91
8-4-15
Handrest Replacement Procedure
NOTE:
If required, the main shelf handrest may be replaced separately, as described below.
8-4-15-1
Tools
Appropriate Phillips screwdrivers.
8-4-15-2
Preparations
Shut down the Vivid q N ultrasound unit.
8-4-15-3
Handrest Removal Procedure
1) Remove all peripherals from the Vivid q N portable ultrasound scanner.
2.) Working from the
front
of the SafeLock Cart and below the main shelf, loosen and remove the three
screws that secure the handrest to the main shelf - see
.
3) Remove the handrest.
8-4-15-4
Handrest Installation Procedure
1) Install a replacement handrest on the SafeLock Cart’s main shelf, making sure the three screw
holes are aligned correctly.
2.) Return and fasten the three screws (previously removed) to secure the handrest in position.
Figure 8-106 Handrest Secured to Main Shelf (viewed from below)
Main Shelf
Handrest
Securing screws
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...