GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-84
Section 8-4 - SafeLock Cart Components Replacement
8-4-11
Gas Spring Replacement Procedure
8-4-11-1
Tools
Appropriate Phillips screwdrivers, Allen key and open wrench.
8-4-11-2
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-4-11-3
Gas Spring Removal Procedure
1) Turn OFF power to the SafeLock Cart and disconnect the AC power cable from the AC input
assembly at the base of the cart (rear).
2.) Remove the Upper Cover (Rear), as described in the
"Upper Cover (Rear) Removal Procedure" on
3) Remove the Lower Cover (Rear), as described in the
"Lower Cover (Rear) Removal Procedure" on
4) Disconnect the Main Harness cable from the AC unit, cut both tie clips and free the Main Cable
Harness so that it does not interfere with the Gas Spring.
5) Gradually raise the Upper Assembly until the lower Gas Spring securing screw is accessible.
6) Lie the SafeLock cart on its left side with the Gas Spring release handle uppermost.
7) Remove the lower Gas Spring securing screw.
8) Remove the Gas Spring by turning it counter-clockwise until it is released from the Gas Spring
Handle assembly.
NOTE:
To remove the Gas Spring handle, unscrew the upper Gas Spring securing screw.
Figure 8-98 SafeLock Cart shown with Gas Spring Handle Uppermost
Gas Spring Handle
Lower Gas Spring securing
screw
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...