Section 8-2 - External Component Replacement Procedures
8-2-5
Rear Cover & Latch Assembly Replacement Procedure
NOTE:
This section describes the replacement procedure for Rear Cover & Latch Assembly - supplied
as part of the Cover Kit (Vivid i n systems Part No. S2423361; Vivid q N systems Part No.
S2423362).
8-2-5-1
Tools
None required.
8-2-5-2
Time Required
30 minutes
8-2-5-3
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-2-5-4
Rear Cover & Latch Assembly Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid q N portable ultrasound scanner.
2.) Remove the Bearing Handle as described in the
"Bearing Handle Removal Procedure" on page 8-2
.
3.) Remove the Control Panel and Keyboard as described in the
Removal Procedure" on page 8-10
4.) Remove the LCD Display Frame as described in the
"LCD Display Frame Removal Procedure" on
5.) Remove the LCD Display.
6.) Loosen and remove the two screws on each of the
left
and
right
hinges (attached to the Rear Cover
and Latch Assembly), as shown in
.
7.) Lift the Rear Cover and Latch Assembly away from the Bottom Assembly and remove completely
from the system.
Figure 8-24 Removing the Hinge Screws
Содержание FQ000040
Страница 2: ......
Страница 10: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL viii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 11: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL ix 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 12: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL x 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 13: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xi 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 14: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL xii 3 SEP 2015 FRONTMATTER FM P R E L I M I N A R Y ...
Страница 348: ...GE DIRECTION FQ091019 REVISION 2 VIVID Q N SERVICE MANUAL 6 4 Figure 6 6 Identifying Monitor Type ...
Страница 697: ......
Страница 698: ... 2013 and 2015 General Electric Company GE Vingmed Ultrasound AS P O Box 141 N 3191 HORTEN NORWAY www gehealthcare com ...