Chapter 8 - Replacement Procedures
8-45
8-3-4
BIOS Battery Replacement
NOTE:
Since the BIOS battery is not available as a spare part (due to shelf life), it should be obtained locally.
The battery model is CR2032.
8-3-4-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-4-2
Time Required
15 minutes
8-3-4-3
Preparations
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-3-4-4
BIOS Battery Removal Procedure
1) Remove the BEP, as described in the
BEP Removal Procedure on page 8 - 38
2) Take out the BIOS battery located on the bottom of the BEP, as shown in
.
8-3-4-5
BIOS Battery Installation Procedure
NOTE:
Before proceeding, refer to the
Preparation Procedure on page 8 - 43
.
1) Fit a new BIOS battery in position on the BEP - refer to
.
2) Re-install the BEP, as described in the
BEP Installation Procedure on page 8 - 42
3.) Turn ON the system and boot-up.
4.) Adjust to the correct date and time settings - see
Date and Time Adjustments on page 3 - 71
5.) Proceed to perform the following functionality test:
-
Battery Parameters Test on page 7 - 102
Figure 8-54 BEP - Location of BIOS Battery